Operating Instructions

4 Operating Instructions

Preparations (Premier Plus)

1.  Fill the medicament bottle with water or medicament solution  

according to the Medicaments that can be used, listed on page 26. 

2.  Screw the bottle connector onto the bottle and push it onto the con-

nector. See picture in section 2.4

3.  Check that the power cord is connected and the unit is in stand-by 

mode, the green indicator light is illuminated.

WARNING

A tip that is bent, altered, or 

worn more than 2 mm will 

lose performance and must 

be exchanged. Prolonged 

use may cause tip breakage 

and injury to the patient.

The operator should be 

aware that ultrasonic  

instruments with small  

diameters are subject to 

breakage at any time. If not 

used correctly or with too 

much power or force the  

instrument WILL break.

Do not use nickel-titanium  

files, since they easily break 

at high frequencies. 

1a.  Attach handpiece cord onto the scaler handpiece. 

1b.  Gently slide the grip onto the scaler handpiece.

2.  Carefully place the tip in the torque wrench.

3.  Use the torque wrench and screw the tip clockwise onto  

the scaler handpiece. Tighten until resistance and the torque 

wrench slides. The torque wrench prevents the tip from being 

overtightened. 

4.1 Scaler

Summary of Contents for Suvi Premier

Page 1: ...Ultrasonic Piezo Scaler Place Disk Owner s Guide...

Page 2: ...53 BioSonic Suvi Premier Plus Medicament feeder AUS Plug Type I 230 60014254 Shaded and italicized items in the table above are not currently offered for sale Important Read this manual carefully befo...

Page 3: ...Premier Premier Plus 3...

Page 4: ...4 2 How to use the scaler 19 4 3 After treatment 19 Automatic cleaning function 19 5 Cleaning and Maintenance 20 5 1 Cleaning of the equipment and components 20 Cleaning 20 Disinfecting 20 Drying 20...

Page 5: ...Premier Premier Plus 5...

Page 6: ...EMC information provided in section 8 2 Portable and mobile RF Radio Frequency communications equipment can affect the product Contraindications Do not use the scaler on patients with cardiac pacemak...

Page 7: ...ning indicates a potentially dangerous situation Non observance may lead to death or injury Caution indicates a potentially harmful situation Non observance may damage the equipment Note indicates a s...

Page 8: ...features of the equipment read Equipment description section 2 6 Caution indicates a potentially harmful situation Non observ ance may damage the equipment Note indicates a situation where special not...

Page 9: ...soft silicone removable grips give the user a comfortable relaxed grip as well as an excellent feel BioSonic Suvi Premier Premier Plus is highly adaptable to any procedure or user approach It is not o...

Page 10: ...POWER OFF key 4 Working mode 1 key 0 to 40 power 5 Working mode 2 key 0 to 70 power 6 Working mode 3 key 0 to 100 power 7 StandBy indicator 8 POWER ON indicator 9 Dry mode indicator 10 Irrigation mod...

Page 11: ...12 OFF position 13 ON position 14 Diagnostic function turn LED light ON or OFF by a single click on the pedal 15 Irrigation position 16 Zero power position 17 Maximum power position 18 Power regulati...

Page 12: ...The medicament bottle can be used for either medicament solutions or ordinary clean water The unit contains an electrically driven air compressor When operating the unit the compressed air forces the...

Page 13: ...unit Caution Consult accompanying documents Type B applied part according to the degree of protection against electrical shock Fuse Input Output This appliance is labeled in accordance with European...

Page 14: ...zed Representative IPX1 Ingress Protection Rating Device has dripping water protection Date of Manufacture Autoclave at 132o C Caution Federal United States laws restricts this device to sale by or on...

Page 15: ...unit in the immediate vicinity of sources of electromagnetic radiation for example the Perfect TCS II or other electrosurgery equipment Connect the foot pedal cable to the foot control pedal and at th...

Page 16: ...t the voltage rating on the rear side matches the voltage of the AC power outlet Verify that the AC power outlet is provided with a protective ground Connect the power cord to the unit and the AC powe...

Page 17: ...ormance and must be exchanged Prolonged use may cause tip breakage and injury to the patient The operator should be aware that ultrasonic instruments with small diameters are subject to breakage at an...

Page 18: ...the ON button activates the dry mode for scaling without water medicament 7 A third press of the ON button activates the irrigation mode The irrigation mode can also be activated in the scaling mode b...

Page 19: ...t a squeaking sound is heard loud and dominating the pressure against the tooth might be too low or the tip not parallel to the surface of the tooth To keep the tip working parallel to the surface of...

Page 20: ...ar broken tip or distortion is noted 3 Clean thoroughly using an ultrasonice cleaner such as the Colt ne Whaledent BioSonic Ultrasonic Cleaning System with BioSonic UC32 Solution Concentrate Disinfect...

Page 21: ...o rinse the lines from cleaning agent solution put clean water in the bottle and run the automatic cleaning cycle until clean water comes out of the handpieces Sterilizing Package in an FDA approved w...

Page 22: ...s 1 Disconnect the power cord from the AC power outlet and the unit 2 Open the fuse holder on the rear side of the unit 3 Inspect the fuses for damage Replace damaged fuses with new ones Verify the fu...

Page 23: ...nd the fuse panel are OK 4 If the problem still remains contact your dealer for support and indicate Error Code E X02 Type of problem Please see A Unit is not responding and no lights are lit on the c...

Page 24: ...ration 1 Check that the tip is firmly tightened Preferably use a torque wrench and tighten clockwise until the torque wrench slides The torque wrench prevents the tip from being over tightened 2 Check...

Page 25: ...y another tip 6 If the problem still remains contact your dealer for support and indicate Error Code E S04 F Tip does not fit smoothly onto the handpiece 1 Clean handpiece threads with compressed air...

Page 26: ...00 g Voltage 100 Vac 50 60 Hz 115 Vac 50 60 Hz 230 Vac 50 60 Hz Primary fuse T500 mAH 250 V 5x20 mm 100 Vac T400 mAH 250 V 5x20 mm 115 Vac T200 mAH 250 V 5x20 mm 230 Vac Power cord Separate with prote...

Page 27: ...angurinara extract Sodium hypochlorite 1 solution WARNING Immedately after using any kind of medicament in the medicament bottle run the automatic cleaning cycle with clean water in the medicament bot...

Page 28: ...nt products are manufactured from new parts or new and serviceable used parts Regardless our warranty terms apply The products are guaranteed in accordance with the terms of the written Certificate of...

Page 29: ...antability and fitness for a particular purpose In no event shall C W be liable for direct indirect tort exemplary punitive special incidental or consequential damages arising from the possession or u...

Page 30: ...output lines 1 kV differential mode 2 kV common mode NOTE U T is the a c mains voltage prior to application of the test level Power frequency magnetic fields should be at levels characteristic of a ty...

Page 31: ...ic immunity Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the BioSoni...

Page 32: ...nt in which radiated RF disturbances are controlled The customer or the user of the equipment can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobil...

Page 33: ...i Grip Color Lilac 4 60018765 Suvi Grip Color Pink 4 60018766 Suvi Grip Color Blue 4 60019328 Subgingival Nozzel 1 60018486 Polisher Nozzle Hooked 1 60018485 Polisher Nozzle Universal 1 60014456 Torqu...

Page 34: ...283 PE 40R General Scaling 60014284 PE 40H General Scaling 60019608 PE 40HS General Scaling 60014285 PE 41PS General Scaling 60014286 IM 1 Scaling Implant maintenance 4 pk 60014287 IM 2 Scaling Implan...

Page 35: ...Premier Premier Plus 35...

Page 36: ...ne Whaledent AG Feldwiesenstrasse 20 9450 Altst tten Switzerland Tel 41 0 71 757 53 00 Fax 41 0 71 757 53 01 info ch coltene com Colt ne Whaledent GmbH Co KG Raiffeisenstrasse 30 89129 Langenau Germa...

Reviews: