4.  Operation 

Recommendation during operation 

 

Using the unit for the first time or after storage, set heating temperature to 100°C and 
preheat the plate during 30 min, it will help to reduce moisture inside the unit. 

4.1. 

Connect the unit to a properly grounded power socket. 

4.2. 

Place  a  glass  or  another  chemical  vessel  with  liquid  on  the  working  plate  and  dip 
magnetic stirrer element in it.  

 

Note!

 

Vessel  must  be  flat-bottomed  and  fit  tightly  to  the 
working surface of the magnetic stirrer. 

4.3. 

Plate temperature control. Using the 

Heating

 knob (fig. 2/2), switch the heating 

on

 

and  set  the  required  temperature  (linearly,  in 

the  30  to  330°C  range).  Heating 

indicator  starts  flashing  (4  times  per  second)  until  the  unit  reaches  the  set 
temperature. 

 

Note!

 

The heating switches off when the plate temperature exceeds the set 
temperature for 

over 30°С.  

 

Caution!

 

Surfaces become hot during operation. Do not touch to avoid burns. 

4.4. 

Using  the 

Mixing

  knob  (fig.  2/1),  switch  the  mixing  mode 

on

  and  set  the  required 

speed. 

4.5. 

After finishing the operation switch 

off

 the 

Mixing

 and 

Heating

 knobs. 

4.6. 

Disconnect the unit from electric circuit. 

Summary of Contents for msh-300

Page 1: ......

Page 2: ...2...

Page 3: ...3 Contents 1 Safety Precautions 4 2 General Information 5 3 Getting started 6 4 Operation 7 6 Specifications 8 7 Maintenance 9 8 Warranty and Claims 10 9 EU Declaration of Conformity 11...

Page 4: ...the serial number label Do not plug the unit into an ungrounded power socket and do not use an ungrounded extension lead Ensure that power switch and plug are easily accessible during use Disconnect...

Page 5: ...talysis field different viscosity chemical reagents dissolving BIOCHEMISTRY solutions preparation dialysis salt and alcohol sedimentation of macromolecules gradient forming in the column chromatograph...

Page 6: ...1 pce Operating manual Certificate 1 copy 3 3 Setup Place the unit upon even horizontal non flammable surface at least 30 cm away from any flammable materials Plug the power cable into the socket on...

Page 7: ...urface of the magnetic stirrer 4 3 Plate temperature control Using the Heating knob fig 2 2 switch the heating on and set the required temperature linearly in the 30 to 330 C range Heating indicator s...

Page 8: ...us stirring time 168 hours 7 days Maximum stirring volume 15 l Operation with magnetic stirring elements of length 10 50 mm Maximum viscosity of stirring liquid 1170 mPa s Plate material aluminium all...

Page 9: ...ry equipment can be used for cleaning and decontamination of the unit 7 4 Improper magnetic stirring elements storage storing several elements together which causes unpredictable magnetic domain disor...

Page 10: ...rranty 8 3 Warranty covers only the units transported in the original package 8 4 If any manufacturing defects are discovered by the Customer an unsatisfactory equipment claim shall be compiled certif...

Page 11: ...rement control and laboratory use EMC requirements General requirements LVS EN 61010 1 2011 Safety requirements for electrical equipment for measurement control and laboratory use General requirements...

Page 12: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 6742 6137 Fax 371 6742 8101 http www biosan lv Edition 4 05 October 2016...

Reviews: