10.7.  Troubleshooting 

Table 5. Troubleshooting 

Symptom 

Possible cause 

Action required 

Manifold dosing error 

Volume of washing 

solution does not 

correspond to the set 

volume, dosing 

unevenness is 

observed along the 

plate wells or the 

washing solution is 

not dispensed at all. 

Poor contact between the 

bottle hose connector and 

the device 

Ensure proper connection of the bottle with the 

device. 

The hose is overbent 

Check if the hose is overbent and straighten it if 

necessary. 

Obstruction of sieve filters 

Check if the sieve filters in bottles No. 1, 2 and 3 

are clogged. 

Obstruction of dispensing 

channel of the manifold or 

a needle 

Check if the dispensing channel of the manifold or 
needles is not obstructed. If yes, perform cleaning 

(see “Obstruction of dispensing channel or a 

needle i

n the manifold” in this Table). 

Absence of liquid in bottles 

Ensure presence of liquid in the bottles No. 1, 2 

and 3; fill up liquid in a bottle if needed. (Attention: 

with 4CHW Logger, information on liquid levels in 

bottles is shown on the display). 

Obstruction of 

dispensing channel 

or a needle in the 

manifold 

During operation needles 

of the manifold can be 

obstructed. 

Determine the obstructed needle by checking the 

underfilling of a well. 

Warning! 

Be careful working with the manifold, 

prevent pricks and injuries caused by needles, they 

may contain dangerous infections! 

Take off the manifold from the device, put it down 

with needles upward. Clean the aspiration channel 

needles with the wiping rod Ø 0.80 mm, and clean 

needles of the dispensing channel with the wiping 

rod Ø 0.45 mm, inserting the wiping rod against 

stop and removing it. 

After cleaning, install the manifold on the device 

and do the rinsing procedure to ensure proper 

operation of the needle. 

In case of repeated obstruction, execute 

requirements of pp. 1 and 2, unscrew the 

manifold’s plugs, disconnect the dispensing and 

aspirating tubes, clean the dispensing or aspiration 
channel with the wiping rod, rinse the manifold with 

the 75% solution of ethanol, replace the plugs, 

install the manifold on the device and ensure the 

proper operation of needles. 

Washing solution is 

not dispensed into a 

plate 

Deformation (wall gluing) 

of dosing valve tube due to 

the long-term storage or 

standstill of the device 

Remove the tube from the valve. 

In the case of deformation, open with fingers the 

closed part of the dosing valve. 

Insert the tube into the valve opening to position 

shown in fig. 2/4. 

Note:

 In order to release the tube, press the valve 

on the side and slightly stretch the tube pulling it 

through the opening. 

Wear or damage of 

the valve tubes and 

the pump 

The connecting tubes are 

worn out or damaged 

during exploitation. 

Turn off the device and have it checked by a 

specialist. Only service engineers and trained 

specialists are allowed to replace the tubes. 

 

 

 

Summary of Contents for IW-8

Page 1: ......

Page 2: ...2...

Page 3: ...Program error messages 16 8 Specifications 16 9 Ordering information 17 10 Maintenance and troubleshooting 18 11 Warranty 21 12 Glossary 22 13 EU Declaration of conformity 23 1 About this edition of...

Page 4: ...rd the results obtained and repeat the washing procedure Accuracy of the obtained testing results depends on the quality of the microplate rinsing procedure To assure correctness of the rinsing proced...

Page 5: ...erflow the maximum level during operation there is a mark on the side surface of the bottle Take necessary precautions utilizing waste liquid in accordance with general laboratory standards 2 4 AFTER...

Page 6: ...W 8 the unit provides automatic control of reagent total volume in all bottles that is necessary for washing and waste collecting and estimates the volume of reagents necessary for washing Microbiolog...

Page 7: ...bes outside inside diam length 6 3 600 mm 5 pieces 7 Tube for manifold outside inside diam length 3 2 1 6 400 mm 1 piece 8 Tube for manifold outside inside diam length 5 3 440 mm 1 piece 9 Tube for hy...

Page 8: ...in a dry dust free place If the unit will be subjected to low temperatures dry the tubes as described in 6 Operation closedown Use the original packaging for transportation whether transported by air...

Page 9: ...4 4 6 Install the manifold in the slots of the holder arm fig 2 3 Magnets in the manifold and the holder arm connect and align the manifold 4 4 7 Connect the manifold inlet connector supplying the li...

Page 10: ...ce bottles near the unit 4 4 19 If 4 channel weight logger is included in the kit remove pads from underside of scale cups Place bottles on the 4 channel weight logger according to the numbering Conne...

Page 11: ..._30 5 1x400_30 Soak 5 min Dispense 200 Dispense NO YES YES YES YES YES YES YES YES Aspirate YES YES YES YES YES YES YES YES YES Shake NO NO NO NO NO NO NO NO NO Dispense rate 02 02 02 02 02 02 02 02 0...

Page 12: ...ttle to use for the main washing Second chan cycles 01 15 Number of washes with second solution 2 channel 1 3 Number of bottle to use for the second washing 5 4 Install the plate on the platform Atten...

Page 13: ...y bottle number using and keys and press the Run Stop key 5 16 During operation display shows program number currently used bottle number and the number of cycles left fig 6 If the current action has...

Page 14: ...section 6 Operation closedown of this manual Attention Always perform the rinsing cycle using distilled water after finishing the operation at the end of working day This will help to keep the unit in...

Page 15: ...l figure 3 6 2 4 Press the Run Stop key fig 4 7 and the system performs a rinsing cycle Repeat the procedure twice if necessary 6 2 5 Repeat 6 2 2 6 2 4 for the remaining tubes 6 3 After finishing the...

Page 16: ...ration in cold rooms and closed laboratory rooms at ambient temperature from 4 C to 40 C in a non condensing atmosphere and maximum relative humidity 80 for temperatures up to 31 C decreasing linearly...

Page 17: ...IW 8 Intelispeed Washer V 4AW BS 060106 AAI 9 2 To inquire about or order the optional accessories or the replacement parts contact Biosan or your local Biosan representative 9 2 1 Optional accessori...

Page 18: ...m the rinsing cycle twice using distilled water after finishing the operation see section 5 10 4 2 Monthly With 75 ethanol solution clean mounting surface of the plate platform and plate holder spring...

Page 19: ...sed by needles they may contain dangerous infections Take off the manifold from the device put it down with needles upward Clean the aspiration channel needles with the wiping rod 0 80 mm and clean ne...

Page 20: ...service engineer 10 8 Storage and transportation 10 8 1 Remove the manifold tubes and platform for plates 10 8 2 Put on packaging blocks onto sides of the unit and place it into the original carton bo...

Page 21: ...ddress To obtain the claim form visit Technical support page on our website at link below 11 5 Extended warranty For IW 8 the Premium class model one year of extended warranty is available free of cha...

Page 22: ...ich provides solution aspiration consequently from two opposite points of a well Strip washing mode plate washing by rows Dispensing filling wells with a set solution volume Waste bottle the bottle fo...

Page 23: ...l equipment for measurement control and laboratory use General requirements EMC 2014 30 EU LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements General...

Page 24: ...Biosan SIA Ratsupites 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 02 November of 2020...

Reviews: