12 

5.3. 

All possible program parameters and their meaning are shown in Table 2.  

Table 2. Program parameters and their definitions 

Parameter 

Value or range 

Description 

Parameter 

Value or range 

Description 

Dispense 

yes/no 

Perform fill 

Aspirate 

yes/no 

Perform aspiration

1

 

Shake 

yes/no 

Plate shaking on the platform during the cycle 

Dispense rate 

01..03 

Speed of liquid filling (100; 200; 300 μl/s) 

Aspirate rate 

01..03 

Speed of liquid aspiration 

Soak limit, s 

0..300 (step 10 s) 

Time between fill and aspiration 

Shake limit, s 

05..150 (step 5 s) 

Shaking time  

Double aspiration 

yes/no 

Second aspiration mode 

Dispensed volume 

25…1600 (step 25) 

Volume of dispensed liquid 

Aspirate time, ms 

200..3000 ms (step 

200 ms) 

Time of aspiration from well in cycle 

Final aspirate, s 

200..3000 ms (step 

200 ms) 

Time of last aspiration in the cycle 

First aspirate 

yes/no 

Aspiration is the first action in a cycle 

Wash by rows 

yes/no 

Each row is washed once per cycle 

On two channels 

yes/no 

Use 2 washing solutions 

Num.of wash cycles 

1..15 

Number of washes with first solution 

1 channel 

1..3 

Number of bottle to use for the main washing 

Second chan. cycles 

01..15 

Number of washes with second solution 

2 channel 

1..3 

Number of bottle to use for the second washing 

5.4. 

Install the plate on the platform. 

 

Attention!

 

When installing a plate the first time or when a different type of plate 
is installed perform plate setup to adjust the manifold lowering depth 
(see paragraph 4.20). 

5.5. 

Use the 

+

 and 

-

 keys (fig. 4/6) to select required washing program from 00 to 100. 

5.6. 

Press  the 

Program  Parameters  (Save)

  key  (fig.  4/1)  to  view  the  parameters  of 

program. In the program, 1st parameter is activated (Dispense...yes/no). 

5.7. 

Use the 

+

 and 

-

 keys to select necessary value and press the 

Enter/Calibr.

 key (fig. 

4/4) to save changes to the active parameter and proceed to the next. This way you 
will be able to review and/or change all parameters.  

5.8. 

Press  the 

Back/Row 

  key  (fig.  4/3)  to  return  to  the  previous  parameter  in  the 

program.  

5.9. 

Press the 

Program Parameters (Save)

 key to exit parameter view and change mode 

and save the program. 

5.10.  Press the 

Esc

 key (fig. 4/5) to exit parameter view and change mode without saving.  

5.11.  Press the 

Run/Stop

 key (fig. 4/7) to start the program. 

 

 

 

1

 

During  aspiration,  the  waste  fluid  bottle  is  detected  automatically 

–  bottle  #4.  The  number  is  not 

specified on the display 

Summary of Contents for IW-8

Page 1: ......

Page 2: ...2...

Page 3: ...Program error messages 16 8 Specifications 16 9 Ordering information 17 10 Maintenance and troubleshooting 18 11 Warranty 21 12 Glossary 22 13 EU Declaration of conformity 23 1 About this edition of...

Page 4: ...rd the results obtained and repeat the washing procedure Accuracy of the obtained testing results depends on the quality of the microplate rinsing procedure To assure correctness of the rinsing proced...

Page 5: ...erflow the maximum level during operation there is a mark on the side surface of the bottle Take necessary precautions utilizing waste liquid in accordance with general laboratory standards 2 4 AFTER...

Page 6: ...W 8 the unit provides automatic control of reagent total volume in all bottles that is necessary for washing and waste collecting and estimates the volume of reagents necessary for washing Microbiolog...

Page 7: ...bes outside inside diam length 6 3 600 mm 5 pieces 7 Tube for manifold outside inside diam length 3 2 1 6 400 mm 1 piece 8 Tube for manifold outside inside diam length 5 3 440 mm 1 piece 9 Tube for hy...

Page 8: ...in a dry dust free place If the unit will be subjected to low temperatures dry the tubes as described in 6 Operation closedown Use the original packaging for transportation whether transported by air...

Page 9: ...4 4 6 Install the manifold in the slots of the holder arm fig 2 3 Magnets in the manifold and the holder arm connect and align the manifold 4 4 7 Connect the manifold inlet connector supplying the li...

Page 10: ...ce bottles near the unit 4 4 19 If 4 channel weight logger is included in the kit remove pads from underside of scale cups Place bottles on the 4 channel weight logger according to the numbering Conne...

Page 11: ..._30 5 1x400_30 Soak 5 min Dispense 200 Dispense NO YES YES YES YES YES YES YES YES Aspirate YES YES YES YES YES YES YES YES YES Shake NO NO NO NO NO NO NO NO NO Dispense rate 02 02 02 02 02 02 02 02 0...

Page 12: ...ttle to use for the main washing Second chan cycles 01 15 Number of washes with second solution 2 channel 1 3 Number of bottle to use for the second washing 5 4 Install the plate on the platform Atten...

Page 13: ...y bottle number using and keys and press the Run Stop key 5 16 During operation display shows program number currently used bottle number and the number of cycles left fig 6 If the current action has...

Page 14: ...section 6 Operation closedown of this manual Attention Always perform the rinsing cycle using distilled water after finishing the operation at the end of working day This will help to keep the unit in...

Page 15: ...l figure 3 6 2 4 Press the Run Stop key fig 4 7 and the system performs a rinsing cycle Repeat the procedure twice if necessary 6 2 5 Repeat 6 2 2 6 2 4 for the remaining tubes 6 3 After finishing the...

Page 16: ...ration in cold rooms and closed laboratory rooms at ambient temperature from 4 C to 40 C in a non condensing atmosphere and maximum relative humidity 80 for temperatures up to 31 C decreasing linearly...

Page 17: ...IW 8 Intelispeed Washer V 4AW BS 060106 AAI 9 2 To inquire about or order the optional accessories or the replacement parts contact Biosan or your local Biosan representative 9 2 1 Optional accessori...

Page 18: ...m the rinsing cycle twice using distilled water after finishing the operation see section 5 10 4 2 Monthly With 75 ethanol solution clean mounting surface of the plate platform and plate holder spring...

Page 19: ...sed by needles they may contain dangerous infections Take off the manifold from the device put it down with needles upward Clean the aspiration channel needles with the wiping rod 0 80 mm and clean ne...

Page 20: ...service engineer 10 8 Storage and transportation 10 8 1 Remove the manifold tubes and platform for plates 10 8 2 Put on packaging blocks onto sides of the unit and place it into the original carton bo...

Page 21: ...ddress To obtain the claim form visit Technical support page on our website at link below 11 5 Extended warranty For IW 8 the Premium class model one year of extended warranty is available free of cha...

Page 22: ...ich provides solution aspiration consequently from two opposite points of a well Strip washing mode plate washing by rows Dispensing filling wells with a set solution volume Waste bottle the bottle fo...

Page 23: ...l equipment for measurement control and laboratory use General requirements EMC 2014 30 EU LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements General...

Page 24: ...Biosan SIA Ratsupites 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 02 November of 2020...

Reviews: