2. Remove battery compartment lid, remove battery, insert new equivalent
battery type. Ensure battery is correctly positioned with the “+” facing up.
Carefully replace the battery compartment lid. Ensure “O” ring (washer)
is in place to maintain water resistance. Properly dispose of the batteries
according to local regulations, keeping them from small children and heat.
5h. Error Messages
Display
Error
Er. 1
Calibration test fails
Er. H
Ambient temperature too high:
When ambient temperature is higher then
40.0°C / 104.0°F.
Er. L
Ambient temperature too low:
When ambient temperature is lower than
10.0°C / 50.0°F.
H
Measured temperature too high:
Displays “H” when measured temperature
is higher than 43.99°C / 111.18°F.
L
Measured temperature too low:
Displays “L” when measured temperature is lower
than 32°C / 89.60°F.
Flat Battery indication:
If the battery icon is shown on the display, the
battery should be replaced immediately.
Blank display:
Check if the battery has been loaded correctly by
checking polarity of the battery.
6. Avoiding Pregnancy with Natural Family Planning
In natural family planning, there are several ways of determining the end of
pre-ovulation fertility and the beginning of post-ovulation infertility.
Professionals teaching natural family planning claim most unplanned
pregnancies occur during pre-ovulation infertility. This booklet does not
include guidelines for natural family planning, since it is beyond the scope of
this booklet.
Your Ovulation digital thermometer only helps you determine when ovulation
has occurred. Calculating infertile periods involves other bodily signs in
addition to temperature change and requires professional guidance when
avoidance of pregnancy is your objective.
NOTE:
This product is not for contraceptive use
°