7.0  MFS Maintenance

The MFS is a rugged instrument that with normal use does not require mainte-

nance.

8.0  Return Authorization

Should the need for service arise, please contact Bios International for a RMA

number prior to returning your MFS for service. You can telephone Bios at 

(800) 663 4977 or (973) 492 8400, or email [email protected].

9.0  Shipment

When shipping the DryCal MFS please ensure that the packaging is adequate to

protect the instrument. When possible the DryCal MFS should be shipped in the

original packaging. Bios International is not responsible for damage that occurs

during shipment. 

5

4

5.0  Operating Instructions

For Use as a Stable Flow Source

1

Turn all pneumatic toggle valves to the off (up) position.

2

Connect the MFS inlet on rear of unit to a clean, dry, oil-free compressed gas 

source (150 PSI max.)

3

Toggle the 1K—2K valve to the on (down) position.

4

Set the First Stage regulator (80—100 PSI.)

5

Adjust the Second Stage regulator (30—60 PSI.)

6

Toggle the 1K—2K valve to the off (up) position.

7

Depending on the flow range selected, connect the flow measurement device 

to the Low Flow Outlet (10—1,000 ccm) or the High Flow Outlet (1,000—50,000

ccm), located on the front of the MFS.

8

Select the toggle switch for your desired flow range and toggle it to on 

(down.) Multiple toggle switches may be turned on simultaneously to obtain a 

cumulative flow. Allow sufficient time for thermal stabilization of the selected 

nozzles(s) (lower flows may take longer to stabilize.)

9

Vary the Second Stage regulator’s pressure to dial in the desired flow.

For Use as a Stable Pressure Source

1

Turn all pneumatic toggle valves to the off (up) position. 

2

Connect the MFS inlet to a clean, dry, oil-free compressed gas source 

(150 PSI max.)

3

Toggle the 1K—2K valve to the on (down) position.

4

Set the First Stage regulator (80—100 PSI.)

5

Vary pressure using the Second Stage regulator to select an appropriate 

pressure for the External outlet.

6

Toggle the 1K—2K valve to the off (up) position.

7

Connect the flow measurement device to the External outlet located on the 

back of the MFS.

8

Select the toggle switch for the External outlet and toggle it to on (down.)