22 

 

Step 16. Check the results

 

 

 

During Calibration 

Check the alignment markers have been placed in the correct position. 

Software will recognize two peaks from Alignment Marker signal. DO NOT use Size 

Marker or DNA sample to do “Calibrate”.

 

 

 

Summary of Contents for Qsep 400

Page 1: ...BiOptic Inc 4F No 108 3 Minquan Rd Xindian Dist New Taipei City 23141 Taiwan R O C www bioptic com tw Revolution Qsep400 Operation Manual Hardware...

Page 2: ...s 4 Packing List 4 Cautions 6 1 Qsep400 Instrument 9 1 1 Qsep400 Front View 9 1 2 Qsep400 Back View 11 1 3 Qsep400 Top View 12 2 Installation Instruction 13 2 1 Environment Requirements 13 2 2 Power C...

Page 3: ...is prohibited without prior written permission of BiOptic Inc except as allowed under the copyright laws Qsep Series and Q Analyzer are registered trademarks of BiOptic Inc Microsoft Windows is regist...

Page 4: ...INSTRUCTIONS BiOptic Inc assumes no liability whatsoever for any personal injury property damage or other loss resulting from not complying or familiar with the manuals or improper operation of the d...

Page 5: ...allation Disc including Q Analyzer for Qsep400 and Qsep Series Operation Manual Note The Software key contains 8GB storage space Q Analyzer for Qsep400 and Qsep Series Operation Manual including Hardw...

Page 6: ...ce it upside down 4 Store the cartridge in the calm shell container Cartridge needs to be stored in a proper environment as the instructions of cartridge unpacking guide to maintain its lifetime and q...

Page 7: ...lease disconnect the power plug 15 Please clean the instrument with dry cloth For other maintenance please contact local distributor 16 Please use the MAINS power cord within adequate rating 17 The co...

Page 8: ...ntuitively The disposable gel cartridge with integrated pre programmed test methods makes capillary electrophoresis experiments no longer a painful procedure that requires well trained operators No mo...

Page 9: ...e high voltage source of the capillary electrophoresis is ON B Main Power switch The blue LED lights up when it is turned on C Operation room Open the onboard computer module A rotation platform inclu...

Page 10: ...10 Figure 1 2 Rotation platform...

Page 11: ...iew A Air pressure gauge Display the value of the pressure operating in Qsep400 B Air flow Inlet Port Connect to an external pump using an 1 8 inches tube C USB Port Connect to a computer D Power Conn...

Page 12: ...12 1 3 Qsep400 Top View Figure 1 4 Qsep400 top view A Cartridge door...

Page 13: ...installation it is recommended to leave approximately 711 mm 28 inches clear space above the counter space or table Allow approximately 76 mm 3 inches of clear space at the rear and to the side of the...

Page 14: ...in the package 100 240V AC 47 63Hz The dimensions are 12 2cm L 5 2cm W 3 5cm H and weight of the instrument is 26Kg Direct control the instrument with onboard computer Connect the air source within th...

Page 15: ...ace Qsep400 instrument on a stable platform and ensure it is away from water and any other high power electronic devices Qsep400 needs to be operated in a clean and well ventilated environment Remove...

Page 16: ...Qsep400 The message box of Purge function check will pop up after connected Follow the instruction and proceed the check before using is recommended User needs to perform the purge function check ever...

Page 17: ...y into the tray holder Step 6 Alignment Marker AM Preparation Add at least 20 l Alignment Marker into the 0 2 ml tube 4 tubes and add 10 l Mineral Oil on the top of Alignment Marker Spin down and make...

Page 18: ...l tube that placed at A1 A12 position must be removed cutoff Step 10 Click Park to move the holder back to the park position and click Close Sample Door Step 11 After unpacking cartridge open the cart...

Page 19: ...se proceed with the following steps The storage and transportation condition may influence the cartridge and cause unstable current If current gray line is unstable during HV check please repeat this...

Page 20: ...lumn and designate the sample locations test method and result name following steps 14 1 to 14 4 Step 14 1 Sample Position Mark the position of sample on the plate and then click OK Step 14 2 Method S...

Page 21: ...Para Set the parameters Baseline Factor Peak Threshold Calculate Calculate Reference marker table based on built in size marker data Create size marker Place size marker at SM row on the holder create...

Page 22: ...the results During Calibration Check the alignment markers have been placed in the correct position Software will recognize two peaks from Alignment Marker signal DO NOT use Size Marker or DNA sample...