PRODUCT SHEET
www.biopac.com
BIOPAC Hardware |
BioNomadix Series
| Page 10 - 26
Updated: 8.18.2020
Table 4: BioNomadix Accelerometer
– See Table 1 for Biopotentials, Table 2 for Dual Transducer, and Table 3 for Combo Pairs
Table 5: Common Specs
Operational Range:
10 meters (line-of-sight) typical in standard laboratory setups. See also:
Operational Range and Characteristics
.
Delay:
Large fixed component (15.6 ms) and small variable component (±0.5 ms rms)
Operating Temp & Humidity:
Temperature: 5-45
C
Humidity: 95% non-condensing
Size & Weight:
Transmitter: (approx.): 6 cm x 4 cm x 2 cm: 54 grams
Receiver: (approx.): 4 cm x 11 cm x 19 cm: 380 grams
Transmitter:
Type: Ultra-low power, 2.4 GHz bi-directional digital RF transmitter
Rate: 2,000 Hz (between transmitter and receiver)
Receiver Power:
Use with an MP Research System or with isolated power supply IPS100C for 3rd-party data acquisition system.
Battery & Charger:
BioNomadix transmitters use an L-ion battery: full charge takes approx. 1 hour to provide maximum operating time.
A battery charger is included with each module pair. See BN-CHRG for charge time and recharge cycle details.
Compliance:
FCC,
, IC,
- FCC Part 15 B - FCC ID: receiver: ZWIBNXR1, transmitter ZWIBNXT1
IC: receiver: 9901A-BNXR1, transmitter: 9901A-BNXT1
BioNomadix
BN-ACCL3
Signal type:
G (X, Y, Z)
Bandlimits
Max:
Factory preset:
Filter Options:
Alternative signal:
±2, ±4, ±8 or ±16 G
± 16 G at 400 Hz LP
DC to 3.13 Hz LP up to 400 Hz LP (in power of 2 steps)
Tap Event Mark Mode (
replaces
G)
Resolution
:
X: 5 mg rms, Y: 6 mg rms, Z: 9 mg (rms) (±2 G scale at 400 Hz LP)
Signal range:
Selectable:
±2, ±4, ±8 or ±16 G
Output Voltage range:
±10 V (receiver output)
Operating time:
72-90 hours
Included strap:
33 cm - BN-STRAP33
Input:
Attach BioNomadix transmitter to subject
– no additional hardware input required; sensor is internal to transmitter.