9. Tightly replace the cap on the Control Solution bottle.
10. Compare your Quality Control Test result to the Control Solution Range printed
on the test strip vial label.
Performing a Quality Control Test
CLOSE
9
PRECAUTION
- Your Control Solution results will not be calculated as part of the average reading but still
can be recalled. The Quality Control Test result will be shown with " " on the screen.
- The suggested environment for the Quality Control Test
is 15 - 40°C.
- Before " " and " " appear, do not touch the Control Solution to the sample entry port. If
you do so, the meter will show " " and " ".
- Do not drip control solution directly into the sample entry port. Doing so may
contaminate or damage the meter.
- For detailed information on the validity period after opening and the storage
environment, please refer to RIGHTEST Control Solution GC570 Insert.
34
Recalling Results and Viewing the Average Result
The RIGHTEST Meter is able to store 500 test results with time and date automatically. If your meter
already has 500 results stored, the newest test result will replace the oldest one. To view results from
memory, turn on the meter without a trip inserted.
1. Press the main button to switch from Time and Date to the Memory screen.
"mem" will be be displayed in the upper right corner of the display. Press the
main button to display the latest result. Subsequent clicks of the main button
will show stored results in historical order. The sequence number and year will
be displayed at the bottom of the screen.
NOTE
The main button is used to cycle through tests with increasing sequence number.
Sequence number "1" is the latest result. "500" is the oldest.