C7SM-1.01
23/79
Service Manual of Cardio 7
2-3. Diagnosis
Problem
Resolution
1. The diagnosis shows LVH.
1. If disease is diagnosed, abnormal ECG
always appears.
2. LVH occurs when the sum of one of V1 or
V2 and one of V5 or V6 is 3.5 or higher.
3. Upgrade to the latest version of the
software.
2. Diagnosis shows MI.
1. If MI is diagnosed, it is likely that the print
button was pressed before the wave
patterns are stabilized.
2. Check HR before printout and print it
again.
3. Upgrade to the latest version of the
software.