![biomotum SPARK User Manual Download Page 16](http://html1.mh-extra.com/html/biomotum/spark/spark_user-manual_2754058016.webp)
●
Once the proper size calf cuff has been selected, open the quick-release
lever on the calf cuff (Figure 8) and slide the cuff onto the carbon-fiber
upright. Be sure that the cuff is facing the same direction as the footplate
(i.e., back of the cuff is aligned with the heel of the footplate).
●
Adjust the height of the calf cuff so that it aligns with the largest part of
the calf musculature. This can be done by having the participant stand on
the footplate and sliding the calf cuff up or down to the appropriate height.
(Figure 9)
●
Tighten the quick-release lever on the calf cuff to secure it in place. Secure
the purple cables by snapping them into the clips on both sides of the cuff
bracket.
●
When the quick-release lever on the cuff is secured, the cuff should not
slide up and down. If the cuff moves with the quick-release lever secured,
refer to the Troubleshooting Guide.
NOTE: Calf cuff height adjustment can also be done when donning the SPARK
(after the footplate is in the shoe and the foot is inserted).
Figure 8. Quick release lever in closed (left) and open (right) positions
REC-184 Version 0
Page 16 of 39