BIOMÉRIEUX

VIDAS

®

ESSENTIALS

Summary of Contents for VIDAS

Page 1: ...BIOM RIEUX VIDAS ESSENTIALS...

Page 2: ...er to warm up for 45 minutes Switch on the printer the screen and the computer Wait for the Windows welcome window to appear and enter the user s name and password After 2 or 3 minutes waiting double...

Page 3: ...label using the barcode reader slowly scan the barcode from top to botton until the code has been read completely then finally then finally The MLE Card is then automatically printed Document only for...

Page 4: ...EST To modify an analysis request for a patient before running the series of analyses select the patient from the Prede ned Section or the Job List make the necessary modi cations click to save the mo...

Page 5: ...the barcode and start over In case of other errors MLE data instruments etc celect them for more details make any necessary corrections and delete them by cliking on Refer toVIDAS user s manual for d...

Page 6: ...est Every 6 Months Cleaning of the reagent strip area Yearly A preventive maintenance has to be done by a bioM rieux representative Performing VIDAS QCV test QCV test INSTRUMENT MENU click select the...

Page 7: ...ompliant result is produced in the same section independent of the position PUT THE SECTION OFFLINE and Call bioM rieux Technical Support For R3 If the result of a particular position is outside the r...