HOW TO PERFORM A CALIBRATION? 

Read the MLE bar code

 

Integrate factory data into the instrument

1

When using a new lot

Calibrate

 

Adjustment of the instrument

to the factory data

2

When using a new lot

+

Every 14 or 28 days

Control

 

Check calibration and ensure that reagent

performance has not been altered

After each calibration

+

When using a new kit

HOW TO SCAN MLE DATA?

 

In the

 Calibration menu,

 select.

Select 

Scan Master Lot

.

 

 

                          then   

 

            fi nally

When the windows pop-up, scan the MLE barcode on the reagent kit label using the 

barcode reader: slowly scan the barcode from top to botton until the code has been 

read completely.

The MLE Card is then automatically printed 

Document only for Training use. Does not replace the User Manual.

MENU OVERVIEW

MAIN MENU

Refer to VIDAS

®

 user’s manual for detailed data

1

1

2

2

3

3

4

4

5

5

Loading menu:

 To prepare and run analysis as soon as SPR

®

, strips ans samples 

are in position.

Calibration menu:

 

To display calibrations and to access various procedure for 

the calibration.

Navigation tree:

 To display the VIDAS

®

 menus list.

Results menu:

 To display and analyse results.

Status bar:

 section status, presence of errors, person currently logged, time.

Summary of Contents for VIDAS

Page 1: ...BIOM RIEUX VIDAS ESSENTIALS...

Page 2: ...er to warm up for 45 minutes Switch on the printer the screen and the computer Wait for the Windows welcome window to appear and enter the user s name and password After 2 or 3 minutes waiting double...

Page 3: ...label using the barcode reader slowly scan the barcode from top to botton until the code has been read completely then finally then finally The MLE Card is then automatically printed Document only for...

Page 4: ...EST To modify an analysis request for a patient before running the series of analyses select the patient from the Prede ned Section or the Job List make the necessary modi cations click to save the mo...

Page 5: ...the barcode and start over In case of other errors MLE data instruments etc celect them for more details make any necessary corrections and delete them by cliking on Refer toVIDAS user s manual for d...

Page 6: ...est Every 6 Months Cleaning of the reagent strip area Yearly A preventive maintenance has to be done by a bioM rieux representative Performing VIDAS QCV test QCV test INSTRUMENT MENU click select the...

Page 7: ...ompliant result is produced in the same section independent of the position PUT THE SECTION OFFLINE and Call bioM rieux Technical Support For R3 If the result of a particular position is outside the r...

Page 8: ......

Page 9: ...r registered trademarks belonging to bioM rieux or one of its subsidiaries or one of its companies Any other name or trademark is to the property of its respective owner bioM rieux S A 673 620 399 RCS...

Reviews: