8.6 Reset procedure
Reset is not an usual operation, but it may be necessary to do so after the update of the software or if the
device is not operating correctly.
To reset your Sampl’air Lite, press simultaneously the
Enter and Escape key
during 5 seconds.
In case of jamming of the device, stopping the sampler is also possible when inserting a stick in the
« forced
stop » hole
that is situated at the bottom of the device. Note that the data you have keyed in and saved in the
memory are not deleted when proceeding like this, it resets only the main PCB.
The following data are reset to default values:
o
Program list (10 min / 10 L)
o
Configuration options
o
language : english
o
backlight : off
NOTE :
After resetting, it is necessary to
resume the complete programming ( list of the
programs, choice of the language, backlight
and administrator code).
Are not reset:
o
serial number of unit
o
calibration data
o
operational counters