Biomega BioClave Operation Manual Download Page 6

2 0

3. Packing Content

Instructions manual

Instructions manual

Quantity

Accessory

No

2

3

4

5

6

7

8

9

10

11

Loading tray

Tray rack

Tray handling tool

Door adjustment tool

Draining hose

Instructions manual

Spare fuse power

Spare fuse for valve (3A)

Door seal

3

1

1

1

2

1

2

2

2

1

1

16L Steam sterilizer

1

Pressure psi

UNWRAPPED

Time (min)

15

0

30

15

0

30

Time (min)

UNWRAPPED

Pressure psi

Spare fuse for mainboard (1A)

(20A)

134°C

121°C

3

Summary of Contents for BioClave

Page 1: ...Operations Manual V 1 0...

Page 2: ...you for choosing the Biomega BioClave steam sterilizer Your steam sterilizer has been CE certified and designed with durability reliability and safety in mind It is your responsibility to install thi...

Page 3: ...LED WATER 6 1 WATER PROPERTIES CHARACTERISTICS 18 2 DIAGRAMS OF THE STERILIZATION PROGRAMS 19 3 ELECTRICAL DRAWING 23 4 HYDRAULIC DRAWING 24 APPENDIX 3 23 APPENDIX Instructions manual Instructions man...

Page 4: ...ols It operates automatically with 134 C and 121 C sterilization temperatures This sterilizer is in compliance with the European Directive 93 42 CEE and it has been produced in accordance with the EN...

Page 5: ...0Hz 3 Nominal power 1600W 120V 1250W 220V 4 Sterilization Temperature 121 C 134 C 5 Main Fuses T20A 250V 6 Capacity of the distilled water tank Approx 2 5L water at level MAX 7 OperatingEnvironment 5...

Page 6: ...tray Tray rack Tray handling tool Door adjustment tool Draining hose Instructions manual Spare fuse power Spare fuse for valve 3A Door seal 3 1 1 1 2 1 2 2 2 1 1 16L Steam sterilizer 1 Pressure psi U...

Page 7: ...Wrapped liquid 0 50 0 50 Ensure that the sterilizer is installed with 2 5in 10cm ventilation space on all sides of the sterilizer and 5 in 20cm on top side The clearance required to open the door is...

Page 8: ...le To cancel or stop the cycle press and hold this button for 3 sec Time Display State State description Ld HE Time in Min PL Ed E1 E9 Do preheating loading Door state vacuuming heating sterilization...

Page 9: ...3 Sample Preperation For the most effective sterilization and to preserve the sample please follow below Arrange the samples of different material on different trays or with at least 2in of space betw...

Page 10: ...pump or the door seal Check the air release valve Make sure you have turned the door handle to the max position or check the door switch Check the water pump and the air release valve Check steam gene...

Page 11: ...ton is pressed again do After a cycle is completed the time display will show Ed Once the pressure has reached 0 PSI on the display the door can be opened and samples can be removed 6 8 Sterilization...

Page 12: ...ank The right valve is used for draining the clean water tank Always be sure that the hose is draining into a vessel with at least 2L capacity 10 5 1 Fully open the door 10 5 2 Remove the door seal ca...

Page 13: ...the program there is only one option 2 This is to adjust the drying time TEMP START DRYING Caution Always ensure that the sample being sterilized can withstand the sterilization time that is being pr...

Page 14: ...back of the sterilizer 9 2 Connect the printer power cord to the socket with the proper electrical specifications 9 3 Power on the sterilizer and press the sel button 9 4 After the sterilization cycle...