Biomega BioClave Operation Manual Download Page 13

1 0

1 3

8 Advanced adjustment settings

8.1 Adjustment of sterilization time and drying time

8.1.1  Hold the UNWRAPPED button for 30seconds. The machine will beep

to signify that you have enetered the advanced settings mode.

8.1.2 The pressure display window shows “0” and the temperature display

window shows  - - -. The timer display windowwill show the pre-
programmes sterilization time parameter.

8.1.3 Select the sterilization program that you wish to adjust. For example,

If you want to adjust the parameter of

Press the

button. Press the button again ,the time display shows “1”.

WRAPPED.

WRAPPED

,

0 means the 121 sterilization time.(The range from 3 to 40 min.)

1 means the 134 sterilization time.(The range from 3 to 40 min.)

2 means the drying time.(The range from 3 to 20 min.)

8.1.4  Press

button, the value of time parameter will plus one.

8.1.5 After finishing the adjustment, press

button to confirm it.

8.1.6 When adjusting the

program, there is only one option,

2. This is to adjust the drying time.

TEMP

START

DRYING

Caution: Always ensure that the sample being sterilized can
withstand the sterilization time that is being programmed.

Instructions manual

Instructions manual

10.3  Cleaning the inner-chamber, weekly

-

Remove the trays and tray rack from the inner-chamber

- Wipe the inner chamber with a soft cloth saturated in distilled water
- Apply the same procedure to the trays and tray rack.

Fig. 2

Fig. 1

Caution:
Never attempt to adjust the door seal when the door is in the
closed position.

10.4 Door adjustment

Below are the instructions for the door adjustment. This should only be
performed if the door is not providing the proper seal or if the door seal has
just been replaced.

10.4.1 Open the door and insert the door adjustment tool in the gap beneath the

plastic cover (Fig. 2).

10.4.2 To tighten the door seal, use this tool to adjust the door adjustment nut

counter clockwise (Fig. 1).

10.4.3  If the door seal is too tight (it is difficult to lock/unlock the door),

you may also adjust the door adjustment nut clockwise to loosen it.

Summary of Contents for BioClave

Page 1: ...Operations Manual V 1 0...

Page 2: ...you for choosing the Biomega BioClave steam sterilizer Your steam sterilizer has been CE certified and designed with durability reliability and safety in mind It is your responsibility to install thi...

Page 3: ...LED WATER 6 1 WATER PROPERTIES CHARACTERISTICS 18 2 DIAGRAMS OF THE STERILIZATION PROGRAMS 19 3 ELECTRICAL DRAWING 23 4 HYDRAULIC DRAWING 24 APPENDIX 3 23 APPENDIX Instructions manual Instructions man...

Page 4: ...ols It operates automatically with 134 C and 121 C sterilization temperatures This sterilizer is in compliance with the European Directive 93 42 CEE and it has been produced in accordance with the EN...

Page 5: ...0Hz 3 Nominal power 1600W 120V 1250W 220V 4 Sterilization Temperature 121 C 134 C 5 Main Fuses T20A 250V 6 Capacity of the distilled water tank Approx 2 5L water at level MAX 7 OperatingEnvironment 5...

Page 6: ...tray Tray rack Tray handling tool Door adjustment tool Draining hose Instructions manual Spare fuse power Spare fuse for valve 3A Door seal 3 1 1 1 2 1 2 2 2 1 1 16L Steam sterilizer 1 Pressure psi U...

Page 7: ...Wrapped liquid 0 50 0 50 Ensure that the sterilizer is installed with 2 5in 10cm ventilation space on all sides of the sterilizer and 5 in 20cm on top side The clearance required to open the door is...

Page 8: ...le To cancel or stop the cycle press and hold this button for 3 sec Time Display State State description Ld HE Time in Min PL Ed E1 E9 Do preheating loading Door state vacuuming heating sterilization...

Page 9: ...3 Sample Preperation For the most effective sterilization and to preserve the sample please follow below Arrange the samples of different material on different trays or with at least 2in of space betw...

Page 10: ...pump or the door seal Check the air release valve Make sure you have turned the door handle to the max position or check the door switch Check the water pump and the air release valve Check steam gene...

Page 11: ...ton is pressed again do After a cycle is completed the time display will show Ed Once the pressure has reached 0 PSI on the display the door can be opened and samples can be removed 6 8 Sterilization...

Page 12: ...ank The right valve is used for draining the clean water tank Always be sure that the hose is draining into a vessel with at least 2L capacity 10 5 1 Fully open the door 10 5 2 Remove the door seal ca...

Page 13: ...the program there is only one option 2 This is to adjust the drying time TEMP START DRYING Caution Always ensure that the sample being sterilized can withstand the sterilization time that is being pr...

Page 14: ...back of the sterilizer 9 2 Connect the printer power cord to the socket with the proper electrical specifications 9 3 Power on the sterilizer and press the sel button 9 4 After the sterilization cycle...