Open the sealed bag and inspect each
canister to verify the presence of an o-ring
properly seated into the lower groove.
Position each canister into the center
section as depicted below.
After installation, verify that the o-ring of
each canister is fully seated into the cup of
the center section.
Remove the moisture pad from the sealed
bag and install it into the lower section of the
breathing chamber as shown below.
WARNING: Failure to install the moisture
pad will result in scrubber flooding and
caused elevated carbon dioxide levels in
the inhalation gas that could lead to
injury or death.
WARNING:
Users
ARE NOT
permitted to
mix versions of the OrbSorb
®
carbon dioxide
scrubber within a BioPak. Scrubber
canisters installed into the BioPak must be
of exactly the same shape and type.
3.6 Breathing Chamber Lid Installation
Verify that the large o-ring of the breathing
chamber is properly lubricated. Install the lid
onto the top of the breathing chamber
making sure to properly align. Secure the lid
into position by locking the eight (8) slide
locks over the studs.