Directives and manufacturer declaration - electromagnetic emissions
The PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C models are designed to be used in the
electromagnetic environment specified below. PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C
customers or users must ensure that their devices are used in such an environment.
Emissions test
Compliance
Electromagnetic environment - directives
RF emissions
CISPR11
Group 1
The PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-
7130130C models only use RF energy for internal functions.
As a result, RF emissions are extremely low and are unlikely
to cause interference in nearby electronic devices.
RF emissions
CISPR11
Class B
The PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-
7130130C models are suitable for use in all buildings,
including domestic and those connected directly to the
public low-voltage electrical power supply network that
supplies power to domestic dwellings.
Harmonic emissions
CEI 61000-3-2
Class B
Voltage fluctuations/
Flicker
CEI 61000-3-3
Compliant
Directives and manufacturer declaration - electromagnetic immunity
The PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C models are designed to be used in the
electromagnetic environment specified below. PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C
customers or users must ensure that their devices are used in such an environment.
Immunity test
Test level
CEI 60601
Level of
compliance
Electromagnetic environment - directives
Electrostatic
discharges
(ESD)
CEI 61000-4-2
± 6 kV on contact
± 8 kV into the air
± 6 kV on contact
± 8 kV into the air
The floors must be made of wood, concrete or
ceramic tiles. If the floors are covered with
synthetic material, relative humidity must be
at least 30%.
Fast
transient bursts
CEI 61000-4-4
± 2 kV for electrical
supply lines
± 1 kV for
input/output lines
± 2 kV for electrical
supply lines
± 1 kV for input/output
lines
The quality of the electrical power supply
network must equate to that in a typical
commercial or hospital environment.
Transient
overvoltages
CEI 61000-4-5
± 1 kV between
phases
± 2 kV between
phase and earth
± 1 kV between phases
± 2 kV between phase
and earth
The quality of the electrical power supply
network must equate to that in a typical
commercial or hospital environment.
Voltage dips, short
outages and voltage
fluctuations in the
electrical power supply
input lines
CEI 61000-4-11
<5% UV
(>95% UV dip) for
0.5 cycles
<40 % UV
(60 % UV dip) for 5
cycles
<70% UV
(30% UV dip) for 25
cycles
<5% UV
(>95% UV dip) for 5
seconds
<5% UV
(>95% UV dip) for 0.5
cycles
<40 % UV
(60 % UV dip) for 5
cycles
<70% UV
(30% UV dip) for 25
cycles
<5% UV
(>95% UV dip) for 5
seconds
The quality of the electrical power supply
network must equate to that in a typical
commercial or hospital environment. If a PRD-
7130100C, PRD-7130110C, PRD-7130120C or
PRD-7130130C user requires continuous
operation during electrical power supply
network outages, we recommend connecting
the PRD-7130100C, PRD-7130110C, PRD-
7130120C or PRD-7130130C to an
uninterruptible power supply or battery.
Magnetic field at the
electrical network
frequency (50/60 Hz)
CEI 61000-4-8
3 A/m
3 A/m
Magnetic fields at the electrical network
frequency must be at levels that are
characteristic of a typical commercial or
hospital environment.
NOTE UV is the voltage of the AC network before application of the test level.
Directives and manufacturer declaration - electromagnetic immunity
The PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C models are designed to be used in the