5 Clinical
5.1 PRESETS
iLase has 11 pre-programmed settings, plus 2 open slots to allow the user to set and store
customized settings.
Preset Name
Tip
Initiation
●
Peak
Power
Average
Power
Pulse
Length
Pulse
Interval
1
Gingivectomy
Yes
3.00W
1.00W
CP1
(0-10ms)
0.20ms
2
Troughing
Yes
2.00W
1.00W
CP2
(1.00ms)
1.00ms
3
Excision
Yes
1.80W
0.90W
CP2
(1.00ms)
1.00ms
4
Frenectomy
Yes
2.00W
1.00W
CP2
(1.00ms)
1.00ms
5
Implant Recovery
Yes
2.40W
1.20W
CP2
(1.00ms)
1.00ms
6
Crown Lengthening
Yes
1.80W
0.90W
CP2
(1.00ms)
1.00ms
7
Perio Pockets
Yes
1.60W
0.80W
CP2
(1.00ms)
1.00ms
8
Exposure of Unerupted
Teeth
Yes
1.80W
0.90W
CP2
(1.00ms)
1.00ms
9
Aphthous Ulcer
No
0.70W
0.70W
CW
10
Hemostasis
No
0.50W
0.50W
CW
11
Endo
(*)
0.10W
0.10W
CW
12
Custom 1
(*)
0.10W
0.10W
CW
13
Custom 2
(*)
0.10W
0.10W
CW
●
Not required for pre-initiated tips
(*)
Minimum defaults provided for user setting of Endodontic Procedures such as Pulpotomy and Pulpotomy
as an adjunct to root canal therapy
►
The first 8 procedures listed are performed in contact mode. Aphthous Ulcer and
Hemostasis require a defocused mode of application.