While in READY mode, you may adjust the
power level
at any time during the procedure by toggling
up (+) or down (-) the selector switch; however, the
pulse mode
may not be changed in READY mode.
You must return to STANDBY to change the pulse mode.
When your procedure is complete, press the selector switch on the iLase to enter STANDBY
mode. The LED on the iLase will turn an amber color. The system is now in STANDBY mode. You
can modify and save new settings for the Procedure by changing power values and pressing and
holding the selector switch for approximately 1.5 seconds while in READY mode.
NOTE: Pressing the finger switch for 60 seconds while firing will cause the laser to stop firing. To
continue firing, release the finger switch and then press again.
4.9 SELECTING A DIFFERENT PROCEDURE
While in Standby mode in the Main Screen, highlight the Procedure Name using the up (+) motion of
the selector switch, and make your selection by pressing the finger switch. The procedure screen will
be displayed.
Scroll up (+) or down (-) to select a new procedure, or select your own Custom setting.
After a new procedure is selected, the system will go to the Tip Initiation and Main screens in READY
mode. Switch to STANDBY mode If you need to continue modifying the preset.
4.10 SELECTING A PULSE MODE
While in Standby mode in the Main Screen, highlight and select the pulse mode icon to access the
Pulse Mode menu; one of three available pulse modes can be chosen from that menu by using the
selector switch.
Figure 4
-
7:
Pu
l
se Mode Se
l
ectio
n
Screen
.
For a pulse that is 0.1ms on, followed by 0.2ms off, select CP1.
For a pulse mode that is 1.0ms on, followed by 1.0ms off, select CP2
.
For continuous wave ouptut (not pulsed), select CW.
Return to the main screen by selecting one of the available pulse modes
.