18
/
19
The
meter
for
single
‐
patient
use
and
the
lancing
devices
may
be
cleaned
5
‐
6
times
per
day
(Test
5
‐
6
times
every
day),
so
the
meter
will
be
cleaned
and
disinfect
about
11000
cycles
that
include
cleanings
plus
disinfections
(1
cycle
=
1
cleaning
+
1
disinfection)
during
its
5
years
life
cycle.
1.
Before
the
test,
the
user
should
clean
and
wash
hands.
If
you
are
a
caregiver,
(not
the
patient),
please
use
disposable
gloves.
2.
After
test,
take
one
wipe
to
clean
the
meter’s
housing,
buttons,
lens,
mark,
usb
cover
and
support
block,
including
its
front
and
back
for
10
seconds
for
cleaning
until
there
is
no
soil
on
the
surface.
Then
disinfect
the
meter’s
housing,
buttons,
lens,
mark,
usb
cover
and
support
block
use
another
wipe
for
2
min,
keep
the
surface
wet
for
2
min.
3.
Take
off
the
glove;
discard
carefully
according
to
your
health
‐
care
provider’s
instruction.
Note:
1.
Each
disinfection
step
requires
a
pre
‐
cleaning
step.
2.
If
the
meter
is
being
operated
by
a
second
person
who
is
providing
testing
assistance
to
the
user,
the
meter
and
lancing
device
shall
be
cleaned
and
disinfected
first.
3.
The
users
should
wash
hands
thoroughly
with
soap
and
water
after
handling
the
meter,
lancing
device,
or
test
strips.
METER
SPECIFICATION
Model
G
‐
427B
(with
“
Bioland
”
test
strip
and
control
solution)
Machine
size
90mm
(L)×54mm
(W)×13mm(H)
Weight
50g
(excluding
batteries)
Measuring
method
Amperometric
technology
using
glucose
oxidase
Measuring
range
20
mg/dL
~600
mg/dL
Blood
Source
Fresh
capillary
whole
blood
Blood
Volume
Around
1.8uL
Memory
448
blood
glucose
tests
Display
LCD
display
with
backlight
Power
source
DC
3V
(AAA×2
alkaline
batteries)
Battery
Life
Approx.
500
normal
tests
Operating
conditions
50
℉
~104
℉
,
Humidity:25%~80%RH
Storage
Condition
(system
:meter
and
test
strips)
39.2
℉
~104
℉
and below 85% relative humidity
Storage
Condition
(meter)
‐
40
℉
~131
℉
,
Humidity
<
80%RH,
700hPa~1060hPa
Use
life
Five
years
Limitations
of
use:
※
The
blood
Monitoring
System
is
not
intended
for
use
on
neonates.
※
The
blood
Monitoring
System
is
not
intended
for
use
on
artery
blood,
neonates’
serum
and
plasma.