3
4. ELECTRICAL
SPECIFICATIONS
AC-adaptor for Charging Stand or direct charging
- Input voltages and main plug according to local requirements
- Output voltage 7.5 VDC, 300 mA
AC-adaptor for Charging Carousel
- Input voltages and main plug according to local requirements
- Output voltage 9 VDC, 1200 mA
Note:
To guarantee optimal charging, do not use any other than the original
AC-adaptor supplied by the manufacturer!
5. CHARGING OF THE
e
LINE PIPETTE
The Biohit
e
LINE Charging Stand or Carousel are intended to be used only
with Biohit
e
LINE electronic pipettes. However, the Biohit
e
LINE also features
direct charging by using the Biohit
e
LINE AC-adaptor supplied with the Biohit
e
LINE Charging Stand.
The Biohit
e
LINE pipettes leave the factory fully
charged. The battery is protected against discharging
during the storage with a plastic protector strip,
which must be removed before operating the pipette.
However, should the battery become discharged
during storage and recharging is required, a battery
symbol will be observed in the bottom right corner of
the display (Fig. 2.).
Note:
If the battery is fully discharged, the display will be empty and there will
be no response from the operating buttons. After a few minutes on charge,
the battery symbol will be displayed and the + and – symbols will continue
to blink until the battery is fully charged.
5.1. Charging with the Biohit
e
LINE Charging Stand Carousel
The
e
LINE Charging Stand and Carousel operates through the metal contacts
situated on the sides of the pipette (Fig. 3A) and in the head of the Charging
Stand/Carousel (Fig. 3B and Fig.3C).
1. Check that the battery protector strip is removed and the battery cover
properly closed.
2. Place the pipette into the Charging Stand/Carousel. Ensure that the metal
contacts on the sides of the pipette are properly
fi
tted to the tracks of the
metal contacts in the top of the Charging Stand/Carousel.
3. When charging, the battery symbol will be displayed with the + and
symbols
fl
ashing. The charging time for completely discharged batteries
is 4 hours.
Fig. 2