Bioer ThermoQ Instructions Manual Download Page 3

Instructions for ThermoQ Metal Bath                                                                                    Important Notes 

 

Important Notes 

 

1.

 

General Practice 

 

Caution:   

Important information is contained in any item and should be carefully read. Failure to 
observe the instruction would result in damage or abnormal function of instrument. 

 

Warning! 

The warning message requires extremely careful operation of a certain step. Failure to 
observe the instruction would result in serious personal injury. 

 

2.

 

Safety 

During  operation,  maintenance  and  repair  of  this  instrument,  the  following  basic  safety  notes  shall  be 
observed.  In  case  of  failure  to  follow  these  measures  or  the  warnings  or  notes  indicated  herein,  the  basic 
protection  provided  by  the  instrument,  its  safety  criteria  of  design  and  manufacture,  and  its  predicted  use 
range would be impaired. 
 

Caution: 

This instrument is for indoor use; 

protection class: IP20

;   

Hangzhou  Bioer  Technology  Co.,  Ltd.  shall  be  irresponsible  for  the  consequence 
resulted from user’s failure to observe the following requirements. 

 

a)

 

Adapter 

The  instrument  is  supplied  by  24VDC  DC  adapter.  Every  instrument  is  equipped  with  adapter  when 

leaving factory. It is strongly recommended to use equipped adapter. 

b)

 

  Dismantle power line on the adapter   

Generally, the power line attached with the machine should be used. If the power line is damaged, it 

must  be  replaced  and  is  not  allowed  to  be  repaired.  Replace  the  power  line  of  the  same  type  and 
specifications. When using this instrument, the power line should not be pressed with anything or placed 
on the site with large traffic.   

c)

 

Keeping apart from the live circuit   

The  operator  shall  not  disassemble  instrument  without  permission.  Only  qualified  professionals  are 

allowed  to  replace  elements  or  adjust  the  machine  inside.  Replacement  of  elements  under  energizing 
state is prohibited.   

d) Placement of instrument 

This instrument shall be fixed in a low RH and few dust place apart from water source (e.g. water pool 

and  water  pipeline)  and  the  room  shall  be  well  ventilated,  and  free  of  corrosive  gas  or  interference  of 
strong magnetic field. The instrument shall not be placed in a wet or dusty location. 

The openings on both sides and at the bottom of the instrument are for ventilation circulation and in 

order to avoid over-heat of the instrument body. They shall not be blocked or covered. When a single set 
of instrument is used, the interval between ventilation opening on the both side of the instrument and its 
nearest object shall not be less than 25cm. Also, don’t use the instrument on loose or soft surface, or the 
ventilation opening at instrument bottom might be blocked 

Excessive  temperature  would  impair  the  performance  or  result  in  failure  of  the  instrument.  This 

instrument shall not be used in location subjected to direct sunshine. The instrument shall be kept away 
from hot gas, stove and all other heat source. 

If  the  instrument  is  to  be  set  aside  for  a  long  time,  the  power  plug  shall  be  withdrawn  and  the 

instrument covered with soft cloth or plastic film to avoid entrance of dust. 

 

Caution:   

In  any  of  the  following  cases,  immediately  cut  off  the  power  supply,  withdraw  the 
power  plug  from  the  power  socket,  contact  the  supplier  bearing  certificates  for 
maintenance:   

 

Liquid drops into inside of the instrument. 

 

The instrument is rained or watered. 

 

The  instrument  works  abnormally,  especially  when  abnormal  sound  or  odor  is 
generated. 

 

The instrument is dropped or its casing is damaged. 

Summary of Contents for ThermoQ

Page 1: ...Instructions for ThermoQ Metal Bath Hangzhou Bioer Technology Co Ltd...

Page 2: ...hts reserved Without prior written consent from Hangzhou Bioer Technology Co Ltd any part of the manual shall not be duplicated reproduced or translated into any other language Document No BYQ63170000...

Page 3: ...le instrument without permission Only qualified professionals are allowed to replace elements or adjust the machine inside Replacement of elements under energizing state is prohibited d Placement of i...

Page 4: ...ice a Warranty Content Hangzhou Bioer Technology Co Ltd is responsible for replacement for instrument with failures caused by material and manufacturing defects within 1 month after goods delivery Han...

Page 5: ...r qualification of acceptance properly fill in the product acceptance sheet and send or fax the copied sheet to the supplier for filing and maintenance After unpacking the packing box and packing mate...

Page 6: ...4 INSTRUCTIONS FOR INDICATOR LIGHTS AND BUTTONS 5 4 1 Instructions For Indicator Lights 5 4 2 Instructions For Buttons 5 5 OPERATION 6 5 1 Examination Before Startup 6 5 2 Start Instrument 6 5 3 Temp...

Page 7: ...ith pressure hot lid which can prevent evaporation and explosion of lid the heating block of the hot lid is convenient for replacement Metal module avoids the sample from pollution The components of b...

Page 8: ...ng Conditions Ambient temperature 10 C 30 C Relative humidity 70 Power supply DC24V 0 5V 90W HB 120W CHB Requirements for adapter AC100 240V 50 60HZ input current 2 5A Caution Before using the instrum...

Page 9: ...ng Range 0 99h59min Program Segment Setting Function Yes Cycle Setting Function Yes Program Auto memory Function Yes QC Report Print Function Yes Max Segments 9 Max Cycles 99 Hot lid Operating Temp Bl...

Page 10: ...board and functions of all buttons and preparation before startup When using the instrument for the first time read the content in the chapter carefully before startup 1 Structure Diagram Hot lid comp...

Page 11: ...tural temperature decline status The green light is hot lid indicator light If the green light is on it shows that hot lid is under normal working status otherwise it shows that hot lid is not working...

Page 12: ...mode PRO Multi point operation mode About 2 3 seconds later the instrument will begin to work temperature display window will show the real time temperature and corresponding indicator light will turn...

Page 13: ...ork 2 0 Other software PDF reader Lowest configuration Processor Pentium R 4 Internal storage 1GBs Hard disc 10GB 2 Install Software Click ThermalQSetup exe in the accompanying disc to install metal b...

Page 14: ...Instructions for ThermoQ Metal Bath Chapter IV Connect Computer 8 c Choose installation position of software click next d Choose installation folder click next...

Page 15: ...f Complete installation click Finish 3 Online Setup 3 1 Installation of USB Drive USB drive program must be installed before online Click the accompanying disc CP210x exe install USB drive program 3...

Page 16: ...ation The above diagram is software operation interface 1 Menu bar 2 Contents of instrument 3 Area for displaying and editing files 4 Area for displaying running status 5 Status bar 4 1 User Managemen...

Page 17: ...tting user name click OK to save it c Delete user Choose the user to be deleted from the list of equipment click delete or choose the option of delete user with right button When deleting user delete...

Page 18: ...e user name the system will prop up whether to override it or not when exporting it If the program document under the user is being used the user cannot be exported 4 2 File Management Include the pro...

Page 19: ...edit or choose the option of edit file with right button After editing click save in menu bar Maximum sections 9 Maximum recycling number 99 Time scope 1min 99h59min Temperature scope 0 100 c rename...

Page 20: ...ment click paste or the option of copy and paste in the menu of right button to paste the program document under the user name If there is the same user name with the copied one the system will prop u...

Page 21: ...h Export file Click export choose the folder of user from the window popping up click OK to export file If there is file with the same user name the system will prop up whether to override it or not w...

Page 22: ...display all instruments a Connect Choose sub item of equipment name in equipment list click connect or the option of instrument connection in right button menu b Disconnect Choose sub item of equipmen...

Page 23: ...e blank and the maximum byte is 12 Instrument name can only include figure and English characters The same instrument name cannot be sued Instrument address cannot be blank and the same instrument can...

Page 24: ...ing connected the instrument cannot be removed g Display all instruments Click all instruments the list of all instruments will appear to display the information of all instruments 4 4 Other managemen...

Page 25: ...nt stop running local file Open the file to be ran click run the dialog box of run file will pop up Choose the instrument to be ran click OK to run the program Meanwhile the instrument will save the n...

Page 26: ...l time temperature on upper right corner of the running status diagram it could display the real time curve C System setup Click system setup display system setup window the status of hot lid and alar...

Page 27: ...er 21 d Check running history Click run log the form of running history will pop up e Export experimental results After the finishing of file running click to run export experimental results on upper...

Page 28: ...y window shows 002 Fault of hot lid sensor The first time Please tune off power then tune on it Happen again Please contact supplier or manufacturer 4 Display abnormal character Fault of chip The firs...

Page 29: ...Appendix 1 Wiring Diagram HB...

Page 30: ...Appendix 2 Wiring Diagram CHB...

Page 31: ...Memorandum...

Page 32: ...ad Gaoxin Binjiang District Hangzhou P C 310053 Tel 0571 87774252 87774307 After sales Service Tel 87774558 Fax 0571 87774553 After sales Service Fax 87774559 E mail instrument bioer com cn Hot line f...