Instructions for ThermoQ Metal Bath Important Notes
Important Notes
1.
General Practice
Caution:
Important information is contained in any item and should be carefully read. Failure to
observe the instruction would result in damage or abnormal function of instrument.
Warning!
The warning message requires extremely careful operation of a certain step. Failure to
observe the instruction would result in serious personal injury.
2.
Safety
During operation, maintenance and repair of this instrument, the following basic safety notes shall be
observed. In case of failure to follow these measures or the warnings or notes indicated herein, the basic
protection provided by the instrument, its safety criteria of design and manufacture, and its predicted use
range would be impaired.
Caution:
This instrument is for indoor use;
protection class: IP20
;
Hangzhou Bioer Technology Co., Ltd. shall be irresponsible for the consequence
resulted from user’s failure to observe the following requirements.
a)
Adapter
The instrument is supplied by 24VDC DC adapter. Every instrument is equipped with adapter when
leaving factory. It is strongly recommended to use equipped adapter.
b)
Dismantle power line on the adapter
Generally, the power line attached with the machine should be used. If the power line is damaged, it
must be replaced and is not allowed to be repaired. Replace the power line of the same type and
specifications. When using this instrument, the power line should not be pressed with anything or placed
on the site with large traffic.
c)
Keeping apart from the live circuit
The operator shall not disassemble instrument without permission. Only qualified professionals are
allowed to replace elements or adjust the machine inside. Replacement of elements under energizing
state is prohibited.
d) Placement of instrument
This instrument shall be fixed in a low RH and few dust place apart from water source (e.g. water pool
and water pipeline) and the room shall be well ventilated, and free of corrosive gas or interference of
strong magnetic field. The instrument shall not be placed in a wet or dusty location.
The openings on both sides and at the bottom of the instrument are for ventilation circulation and in
order to avoid over-heat of the instrument body. They shall not be blocked or covered. When a single set
of instrument is used, the interval between ventilation opening on the both side of the instrument and its
nearest object shall not be less than 25cm. Also, don’t use the instrument on loose or soft surface, or the
ventilation opening at instrument bottom might be blocked
Excessive temperature would impair the performance or result in failure of the instrument. This
instrument shall not be used in location subjected to direct sunshine. The instrument shall be kept away
from hot gas, stove and all other heat source.
If the instrument is to be set aside for a long time, the power plug shall be withdrawn and the
instrument covered with soft cloth or plastic film to avoid entrance of dust.
Caution:
In any of the following cases, immediately cut off the power supply, withdraw the
power plug from the power socket, contact the supplier bearing certificates for
maintenance:
Liquid drops into inside of the instrument.
The instrument is rained or watered.
The instrument works abnormally, especially when abnormal sound or odor is
generated.
The instrument is dropped or its casing is damaged.