Bioer ThermoQ Instructions Manual Download Page 20

 

Instructions for ThermoQ Metal Bath                                                               

Chapter IV Connect Computer 

— 

14

 

 

 

 
d. Delete file   

Choose the file to be deleted from the list of equipment, click “delete”

.   

If the file is being edited, it cannot be deleted;   

 

e. Copy   

Choose the file to be copied, click “copy”

  or the option of “copy” in the menu of right button to 

copy the file.   

If the program document under the user is being edited, it cannot be copied.   
 
f. Paste 

Choose the option of equipment from the list of equipment, click “paste”

, or the option of “copy 

and paste” in the menu of right button to paste the program document under the user name.   

If there is the same user name with the copied one, the system will prop up whether to override it or 

not when pasting it.   

I f the program document under the existing user name is being used, the pasting cannot be carried 

out;   

Drag and paste with mouse 
 
g. Import file 

Click “import”

, choose the folder of user from the window popping up, click “OK” to import file.   

If there is the same user name, the system will prop up whether to override it or not when importing it.   
If the file with the same name is being used, it cannot be imported.   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for ThermoQ

Page 1: ...Instructions for ThermoQ Metal Bath Hangzhou Bioer Technology Co Ltd...

Page 2: ...hts reserved Without prior written consent from Hangzhou Bioer Technology Co Ltd any part of the manual shall not be duplicated reproduced or translated into any other language Document No BYQ63170000...

Page 3: ...le instrument without permission Only qualified professionals are allowed to replace elements or adjust the machine inside Replacement of elements under energizing state is prohibited d Placement of i...

Page 4: ...ice a Warranty Content Hangzhou Bioer Technology Co Ltd is responsible for replacement for instrument with failures caused by material and manufacturing defects within 1 month after goods delivery Han...

Page 5: ...r qualification of acceptance properly fill in the product acceptance sheet and send or fax the copied sheet to the supplier for filing and maintenance After unpacking the packing box and packing mate...

Page 6: ...4 INSTRUCTIONS FOR INDICATOR LIGHTS AND BUTTONS 5 4 1 Instructions For Indicator Lights 5 4 2 Instructions For Buttons 5 5 OPERATION 6 5 1 Examination Before Startup 6 5 2 Start Instrument 6 5 3 Temp...

Page 7: ...ith pressure hot lid which can prevent evaporation and explosion of lid the heating block of the hot lid is convenient for replacement Metal module avoids the sample from pollution The components of b...

Page 8: ...ng Conditions Ambient temperature 10 C 30 C Relative humidity 70 Power supply DC24V 0 5V 90W HB 120W CHB Requirements for adapter AC100 240V 50 60HZ input current 2 5A Caution Before using the instrum...

Page 9: ...ng Range 0 99h59min Program Segment Setting Function Yes Cycle Setting Function Yes Program Auto memory Function Yes QC Report Print Function Yes Max Segments 9 Max Cycles 99 Hot lid Operating Temp Bl...

Page 10: ...board and functions of all buttons and preparation before startup When using the instrument for the first time read the content in the chapter carefully before startup 1 Structure Diagram Hot lid comp...

Page 11: ...tural temperature decline status The green light is hot lid indicator light If the green light is on it shows that hot lid is under normal working status otherwise it shows that hot lid is not working...

Page 12: ...mode PRO Multi point operation mode About 2 3 seconds later the instrument will begin to work temperature display window will show the real time temperature and corresponding indicator light will turn...

Page 13: ...ork 2 0 Other software PDF reader Lowest configuration Processor Pentium R 4 Internal storage 1GBs Hard disc 10GB 2 Install Software Click ThermalQSetup exe in the accompanying disc to install metal b...

Page 14: ...Instructions for ThermoQ Metal Bath Chapter IV Connect Computer 8 c Choose installation position of software click next d Choose installation folder click next...

Page 15: ...f Complete installation click Finish 3 Online Setup 3 1 Installation of USB Drive USB drive program must be installed before online Click the accompanying disc CP210x exe install USB drive program 3...

Page 16: ...ation The above diagram is software operation interface 1 Menu bar 2 Contents of instrument 3 Area for displaying and editing files 4 Area for displaying running status 5 Status bar 4 1 User Managemen...

Page 17: ...tting user name click OK to save it c Delete user Choose the user to be deleted from the list of equipment click delete or choose the option of delete user with right button When deleting user delete...

Page 18: ...e user name the system will prop up whether to override it or not when exporting it If the program document under the user is being used the user cannot be exported 4 2 File Management Include the pro...

Page 19: ...edit or choose the option of edit file with right button After editing click save in menu bar Maximum sections 9 Maximum recycling number 99 Time scope 1min 99h59min Temperature scope 0 100 c rename...

Page 20: ...ment click paste or the option of copy and paste in the menu of right button to paste the program document under the user name If there is the same user name with the copied one the system will prop u...

Page 21: ...h Export file Click export choose the folder of user from the window popping up click OK to export file If there is file with the same user name the system will prop up whether to override it or not w...

Page 22: ...display all instruments a Connect Choose sub item of equipment name in equipment list click connect or the option of instrument connection in right button menu b Disconnect Choose sub item of equipmen...

Page 23: ...e blank and the maximum byte is 12 Instrument name can only include figure and English characters The same instrument name cannot be sued Instrument address cannot be blank and the same instrument can...

Page 24: ...ing connected the instrument cannot be removed g Display all instruments Click all instruments the list of all instruments will appear to display the information of all instruments 4 4 Other managemen...

Page 25: ...nt stop running local file Open the file to be ran click run the dialog box of run file will pop up Choose the instrument to be ran click OK to run the program Meanwhile the instrument will save the n...

Page 26: ...l time temperature on upper right corner of the running status diagram it could display the real time curve C System setup Click system setup display system setup window the status of hot lid and alar...

Page 27: ...er 21 d Check running history Click run log the form of running history will pop up e Export experimental results After the finishing of file running click to run export experimental results on upper...

Page 28: ...y window shows 002 Fault of hot lid sensor The first time Please tune off power then tune on it Happen again Please contact supplier or manufacturer 4 Display abnormal character Fault of chip The firs...

Page 29: ...Appendix 1 Wiring Diagram HB...

Page 30: ...Appendix 2 Wiring Diagram CHB...

Page 31: ...Memorandum...

Page 32: ...ad Gaoxin Binjiang District Hangzhou P C 310053 Tel 0571 87774252 87774307 After sales Service Tel 87774558 Fax 0571 87774553 After sales Service Fax 87774559 E mail instrument bioer com cn Hot line f...

Reviews: