16 Biodex Medical Systems, Inc. © 2018
Tablet:
1.
Follow the Assembly Guide that was supplied with the Atomlab 500 Tablet for the proper
order to make all connections.
2.
Run the power cable down the support tube and over the back wall of the cabinet.
3.
Fasten both the chamber signal cable and the power cord at the base and along the support
tube with the supplied tie wraps.
4.
Power up the Atomlab 500 Tablet following the assembly guide supplied with the tablet.
Using the Sharps Container Shield
1.
The Sharps Container Shield is designed for use with the 039-413 Sharps Container.
2.
To deposit sharps, lift the hinged cover and release the syringe into the container. Close the
hinged cover.
3.
To change the sharps container:
a.
Unlock the Sharps Container Shield top door.
b.
Grasp the two handles and rotate the top counterclockwise to lift it off and reveal the
container.
c.
Remove the container and dispose appropriately.
d.
Replace it with a new sharps container.
e.
Replace the top door and rotate clockwise to secure.
Leveling the PET Cabinet
The PET cabinet has front leveling pads inside the outer edges of the base channels. The
cabinet can be raised or lowered by loosening the 9/16" Hex Jam Nut on the top of the plate,
turning the 5/8" Hex Nut on the leveling pad clockwise or counterclockwise to level cabinet.
Lock the 9/16" Hex Nut against the plate after the adjustment has been completed.
Figure 2.10. Leveling Pad Adjustment.