View: The View option allows you to view in different modes.
•
Window: Allows cascading, horizontal or vertical placement of windows.
•
Operations: Allows access to different operating modes.
•
Help: Provides help on specific system topics.
The computer operation is a standard Windows format and the Windows format should be used for
setting date and time.
SYSTEM SOFTWARE SET UP
(See Figure 2-6.)
The System Setup option allows the user to enter site information including the facility name, address
and phone number. It also allows editing of the physician and technologist database, and allows the
system to be set for Simulation or Active mode.
To access system setup, click on <File> in the menu bar, then click on <System Setup>. At this screen, a
check in the box preceding any option means that the option is active. To activate or deactivate any
option, click the mouse on the appropriate option box. Setup options are described below.
MCA Interface
•
Simulation Mode: In Simulation mode, the Atomlab 950 will provide mock measuring without
the need for actual activity. This mode allows the user to explore the various screens and
modes of the system.
NOTE: EXITING SIMULATION MODE
When the Atomlab 950 System is turned ON and the MCA box is OFF, the computer prompts you to select
Simulation mode if desired. If you select Simulation mode, the system returns to the desktop. You can then click on
<Start>, followed by <Programs>, <Biodex Medical Systems>, and <Atomlab 950 Simulation> to enter Simulation
mode.
To exit Simulation mode, click on the <X> in the upper right corner of the screen to return to the desktop. You can
then turn the MCA box ON and double-click on the Atomlab 950 desktop icon to access the Atomlab 950 Active mode.
•
Port: The port listed is for the MCA box connection. The standard configuration has the MCA
box connected to COM 2. The small arrow in the box allows the system to function as a
standard pull-down menu.
•
Probe or Well: If your system has only one detector, there should be a check mark next to the
detector only. This will make the detector your only choice on the menu. If you add a detector
at a later time, you will need to access system setup and put a check next to the appropriate
detector to make it available for use.
I-131 MIRD "S" Value
This is preset for 0.022 rads per microcurie hours. Enter any desired change and click on <Apply> The
changes will cause you to restart the system before they can be applied. If you don't want to make any
changes, click on the <Cancel>.
Thyroid Uptake Report
Use the arrow in the box to choose between Efficiency Half-Life and Biological Half-Life.
Site Information
Enter here your facility name and address. The facility name entered here will appear on all printed
reports.
X
2-13
SETUP INSTRUCTIONS
Summary of Contents for 187-130
Page 8: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 2 Figure 2 1 The MCA Uptake Stand ...
Page 10: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 4 Figure 2 2 The MCA Uptake Stand side view ...
Page 20: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 14 Figure 2 6 The System Setup Screen ...
Page 25: ...1 INTRODUCTION 2 19 SETUP INSTRUCTIONS Figure 2 8 The Database Manager Screen ...
Page 28: ...CONTENTS 3 3 ADMINISTRATION Figure 3 1 The Calibration Screen ...
Page 29: ...1 INTRODUCTION Figure 3 2 A Calibration Report ADMINISTRATION 3 4 ...
Page 31: ...1 INTRODUCTION Figure 3 3 The High Voltage Adjustment Screen ADMINISTRATION 3 6 ...
Page 32: ...1 INTRODUCTION Figure 3 4 The High Voltage Report 3 7 ADMINISTRATION ...
Page 34: ...1 INTRODUCTION Figure 3 5 The Chi Square Screen 3 9 ADMINISTRATION ...
Page 35: ...1 INTRODUCTION Figure 3 6 The Chi Square Report ADMINISTRATION 3 10 ...
Page 37: ...1 INTRODUCTION Figure 3 7 An Administration Report ADMINISTRATION 3 12 ...
Page 39: ...ADMINISTRATION 3 14 1 INTRODUCTION Figure 3 8 The Isotope Editing Screen ...
Page 40: ...1 INTRODUCTION 3 15 ADMINISTRATION Figure 3 9 The Select Isotope Window ...
Page 42: ...1 INTRODUCTION 3 17 ADMINISTRATION Figure 3 10 The Detector Efficiency Screen ...
Page 44: ...1 INTRODUCTION Figure 3 11 The Geometric Efficiency Window 3 19 ADMINISTRATION ...
Page 48: ...1 INTRODUCTION 3 23 ADMINISTRATION Figure 3 13 A Spectrum Analysis Report ...
Page 50: ...1 INTRODUCTION Figure 4 1 The Procedure Definition Screen THYROID UPTAKE 4 2 ...
Page 51: ...1 INTRODUCTION 4 3 THYROID UPTAKE Figure 4 2 The Patient Definition Screen ...
Page 56: ...THYROID UPTAKE 4 8 1 INTRODUCTION Figure 4 3 The Thyroid Uptake Screen ...
Page 58: ...THYROID UPTAKE 4 10 1 INTRODUCTION Figure 4 4 A Thyroid Uptake Report ...
Page 59: ...1 INTRODUCTION Figure 4 5 A Time Activity Report 4 11 THYROID UPTAKE ...
Page 63: ...MANUAL MODE 5 2 1 INTRODUCTION Figure 5 1 The Manual Mode Screen ...
Page 64: ...1 INTRODUCTION 5 3 MANUAL MODE Figure 5 2 A Manual Mode Report ...
Page 66: ...1 INTRODUCTION WIPE TEST 6 2 Figure 6 1 The Wipe Counting Screen ...
Page 67: ...6 3 WIPE TEST CONTENTS Figure 6 2 The Wipe Area Setup Screen ...
Page 68: ...CONTENTS WIPE TEST 6 4 Figure 6 3 The Wipe Preferences Window ...
Page 70: ...1 INTRODUCTION Figure 6 4 The Background LLD Screen WIPE TEST 6 6 ...
Page 72: ...1 INTRODUCTION WIPE TEST 6 8 Figure 6 5 The Wipe Results Report ...
Page 76: ...1 INTRODUCTION 7 3 BIOASSAY Figure 7 2 The Bioassay Preferences Screen ...
Page 78: ...7 5 BIOASSAY CONTENTS Figure 7 3 An Individual Bioassay Report ...
Page 79: ...CONTENTS Figure 7 4 An Employee Bioassay Summary BIOASSAY 7 6 ...
Page 80: ...1 INTRODUCTION 7 7 BIOASSAY Figure 7 5 An Individual Bioassay History Summary ...
Page 82: ...1 INTRODUCTION SCHILLING TESTS 8 2 Figure 8 1 Defining a Facility Standard Schilling Test ...
Page 83: ...CONTENTS 8 3 SCHILLING TESTS Figure 8 2 The Schillings Counting Page ...
Page 84: ...CONTENTS SCHILLING TESTS 8 4 Figure 8 3 A sample Schillings Mallinckrodt Report ...
Page 87: ...CONTENTS 8 7 SCHILLING TESTS Figure 8 4 The Diopac Schillings Counting Page ...
Page 88: ...SCHILLING TESTS 8 8 CONTENTS Figure 8 5 A Sample Diopac Schillings Report ...
Page 90: ...CONTENTS HEMATOLOGY 9 2 9 1 The IHSA Counting Page ...
Page 92: ...1 INTRODUCTION HEMATOLOGY 9 4 Figure 9 2 A Sample IHSA I 125 Test Report ...
Page 93: ...CONTENTS 9 5 HEMATOLOGY Figure 9 3 The CR 51 Counting Page ...
Page 95: ...CONTENTS 9 7 HEMATOLOGY Figure 9 4 A Sample Cr 51 Volume Test Report ...
Page 96: ...CONTENTS Figure 9 5 The Cr 51 Red Blood Cell Survival Counting Page HEMATOLOGY 9 8 ...
Page 98: ...CONTENTS HEMATOLOGY 9 10 Figure 9 6 A sample Cr 51 Survival Test Report ...
Page 100: ...CONTENTS Figure 9 7 The GFR Counting Page HEMATOLOGY 9 12 ...
Page 103: ...1 INTRODUCTION 9 15 HEMATOLOGY Figure 9 8 A One Sample Glomerular Filtration Rate Report ...
Page 104: ...1 INTRODUCTION HEMATOLOGY 9 16 Figure 9 9 A Two Sample Glomerular Filtration Rate Report ...
Page 106: ...1 INTRODUCTION HEMATOLOGY 9 18 Figure 9 10 The ERPF Counting Page ...
Page 108: ...1 INTRODUCTION HEMATOLOGY 9 20 Figure 9 10 The ERPF Counting Page ...
Page 109: ...1 INTRODUCTION 9 21 HEMATOLOGY Figure 9 11 An Effective Renal Plasma Flow Report ...