
SYSTEM SHUT-DOWN
The Atomlab 950 system is now fully set up and ready for operation. Before it can be moved, however, it
must be shut down using the following procedure.
1.
At the Windows Desktop screen, which should now be displayed, move the trackball pointer over
the “Atomlab950” icon and double-click the mouse. The About Atomlab950 screen should now be
displayed. A calibration status box is located on the right side of this screen.
2.
Look below the version number of the Atomlab software (under the Biodex Address on the left side
of the screen) and check the detector status. If the cables are connected properly at this point, there
will be a message reading "Connected To MCA."
NOTE: If the message reads "Device Not Found," reconnect the cables leading to the MCA and check all other
connections. If the system will not connect to the MCA, click on <No> and then click on <Yes> to advance to
Simulation mode. You can then exit Simulation mode by clicking on <File> at the top of the screen, selecting
<System Setup>, and then clicking on <Simulation Mode> to remove the check mark. See "System Software
Set Up."
3.
If the detector status is okay, click on <OK> to advance to the MCA Operation screen.
4.
Move the pointer over the <X> at the top right corner of the screen and click the left trackball button
to exit. The Windows Desktop should now be displayed.
5.
Select <Start>. Now click on <Shut Down> and then <OK>. The system will now take several
seconds to shut down the computer. When finished, a message will appear on the screen noting that
it is safe to turn the system OFF. Turn the power strip off at this time. You can now unplug the
system and move it to the location where it will be used.
NOTE: The push handle can be used as a wire wrap to hold the uptake stand power cord during transport.
GENERAL CLEANING INSTRUCTIONS
As required, wipe down the exterior of the unit using a soft rag slightly dampened with alcohol.
POWER-UP AND SELF-TEST
(See Figures 2-3 and 2-4.)
At this point, your Atomlab 950 Medical Spectrometer should be fully ready for operation. It is
recommended that the system be turned ON at least one hour before use or, if possible, left on at all
times. This will provide optimum performance and will not effect the longevity of any part or component
except the monitor, which should be turned OFF if the Atomlab 950 will not be used for a prolonged
period of time.
To Power-Up And Run The Self-Test
Turn on all components, including the printer. The system may immediately advance to the Atomlab 950
program. If not, double-click on the <Atomlab950> icon from the system desktop. Once the Atomlab 950
program is activated, the system immediately performs a self-test. The About Atomlab 950 screen is then
displayed.
NOTE: If the system is in Simulation Mode, see System Setup later in this manual to return to normal operating
mode.
1. INTRODUCTION
2-7
SETUP INSTRUCTIONS
Summary of Contents for 187-130
Page 8: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 2 Figure 2 1 The MCA Uptake Stand ...
Page 10: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 4 Figure 2 2 The MCA Uptake Stand side view ...
Page 20: ...1 INTRODUCTION SETUP INSTRUCTIONS 2 14 Figure 2 6 The System Setup Screen ...
Page 25: ...1 INTRODUCTION 2 19 SETUP INSTRUCTIONS Figure 2 8 The Database Manager Screen ...
Page 28: ...CONTENTS 3 3 ADMINISTRATION Figure 3 1 The Calibration Screen ...
Page 29: ...1 INTRODUCTION Figure 3 2 A Calibration Report ADMINISTRATION 3 4 ...
Page 31: ...1 INTRODUCTION Figure 3 3 The High Voltage Adjustment Screen ADMINISTRATION 3 6 ...
Page 32: ...1 INTRODUCTION Figure 3 4 The High Voltage Report 3 7 ADMINISTRATION ...
Page 34: ...1 INTRODUCTION Figure 3 5 The Chi Square Screen 3 9 ADMINISTRATION ...
Page 35: ...1 INTRODUCTION Figure 3 6 The Chi Square Report ADMINISTRATION 3 10 ...
Page 37: ...1 INTRODUCTION Figure 3 7 An Administration Report ADMINISTRATION 3 12 ...
Page 39: ...ADMINISTRATION 3 14 1 INTRODUCTION Figure 3 8 The Isotope Editing Screen ...
Page 40: ...1 INTRODUCTION 3 15 ADMINISTRATION Figure 3 9 The Select Isotope Window ...
Page 42: ...1 INTRODUCTION 3 17 ADMINISTRATION Figure 3 10 The Detector Efficiency Screen ...
Page 44: ...1 INTRODUCTION Figure 3 11 The Geometric Efficiency Window 3 19 ADMINISTRATION ...
Page 48: ...1 INTRODUCTION 3 23 ADMINISTRATION Figure 3 13 A Spectrum Analysis Report ...
Page 50: ...1 INTRODUCTION Figure 4 1 The Procedure Definition Screen THYROID UPTAKE 4 2 ...
Page 51: ...1 INTRODUCTION 4 3 THYROID UPTAKE Figure 4 2 The Patient Definition Screen ...
Page 56: ...THYROID UPTAKE 4 8 1 INTRODUCTION Figure 4 3 The Thyroid Uptake Screen ...
Page 58: ...THYROID UPTAKE 4 10 1 INTRODUCTION Figure 4 4 A Thyroid Uptake Report ...
Page 59: ...1 INTRODUCTION Figure 4 5 A Time Activity Report 4 11 THYROID UPTAKE ...
Page 63: ...MANUAL MODE 5 2 1 INTRODUCTION Figure 5 1 The Manual Mode Screen ...
Page 64: ...1 INTRODUCTION 5 3 MANUAL MODE Figure 5 2 A Manual Mode Report ...
Page 66: ...1 INTRODUCTION WIPE TEST 6 2 Figure 6 1 The Wipe Counting Screen ...
Page 67: ...6 3 WIPE TEST CONTENTS Figure 6 2 The Wipe Area Setup Screen ...
Page 68: ...CONTENTS WIPE TEST 6 4 Figure 6 3 The Wipe Preferences Window ...
Page 70: ...1 INTRODUCTION Figure 6 4 The Background LLD Screen WIPE TEST 6 6 ...
Page 72: ...1 INTRODUCTION WIPE TEST 6 8 Figure 6 5 The Wipe Results Report ...
Page 76: ...1 INTRODUCTION 7 3 BIOASSAY Figure 7 2 The Bioassay Preferences Screen ...
Page 78: ...7 5 BIOASSAY CONTENTS Figure 7 3 An Individual Bioassay Report ...
Page 79: ...CONTENTS Figure 7 4 An Employee Bioassay Summary BIOASSAY 7 6 ...
Page 80: ...1 INTRODUCTION 7 7 BIOASSAY Figure 7 5 An Individual Bioassay History Summary ...
Page 82: ...1 INTRODUCTION SCHILLING TESTS 8 2 Figure 8 1 Defining a Facility Standard Schilling Test ...
Page 83: ...CONTENTS 8 3 SCHILLING TESTS Figure 8 2 The Schillings Counting Page ...
Page 84: ...CONTENTS SCHILLING TESTS 8 4 Figure 8 3 A sample Schillings Mallinckrodt Report ...
Page 87: ...CONTENTS 8 7 SCHILLING TESTS Figure 8 4 The Diopac Schillings Counting Page ...
Page 88: ...SCHILLING TESTS 8 8 CONTENTS Figure 8 5 A Sample Diopac Schillings Report ...
Page 90: ...CONTENTS HEMATOLOGY 9 2 9 1 The IHSA Counting Page ...
Page 92: ...1 INTRODUCTION HEMATOLOGY 9 4 Figure 9 2 A Sample IHSA I 125 Test Report ...
Page 93: ...CONTENTS 9 5 HEMATOLOGY Figure 9 3 The CR 51 Counting Page ...
Page 95: ...CONTENTS 9 7 HEMATOLOGY Figure 9 4 A Sample Cr 51 Volume Test Report ...
Page 96: ...CONTENTS Figure 9 5 The Cr 51 Red Blood Cell Survival Counting Page HEMATOLOGY 9 8 ...
Page 98: ...CONTENTS HEMATOLOGY 9 10 Figure 9 6 A sample Cr 51 Survival Test Report ...
Page 100: ...CONTENTS Figure 9 7 The GFR Counting Page HEMATOLOGY 9 12 ...
Page 103: ...1 INTRODUCTION 9 15 HEMATOLOGY Figure 9 8 A One Sample Glomerular Filtration Rate Report ...
Page 104: ...1 INTRODUCTION HEMATOLOGY 9 16 Figure 9 9 A Two Sample Glomerular Filtration Rate Report ...
Page 106: ...1 INTRODUCTION HEMATOLOGY 9 18 Figure 9 10 The ERPF Counting Page ...
Page 108: ...1 INTRODUCTION HEMATOLOGY 9 20 Figure 9 10 The ERPF Counting Page ...
Page 109: ...1 INTRODUCTION 9 21 HEMATOLOGY Figure 9 11 An Effective Renal Plasma Flow Report ...