Page 33 / 41
IFU_MAL_EN_V6
INDICATION
CAUSE
TROUBLESHOOTING STEPS
Calibration error
①
Select
“
Error
”
②
Perform the calibration as indicated in the
Mallya
device instructions for use (see
section 5.2.2.
)
Mallya is maybe too
far away from your
smartphone and
therefore
disconnected
Or
①
Move your Mallya device and your
smartphone closer to one another.
A communication
error occurred
①
Select
“
Disconnected
”
②
Reset Mallya as explained in the Mallya
device instructions for use (see
section 9.3.
)
Bluetooth issue or
signal lost
①
Check that the Mallya device is powered
ON.
②
Check that the Mallya device is connected
to the App.
③
Check the Bluetooth connection status on
your mobile phone.
If the problem is not solved, please contact a
BIOCORP PRODUCTION authorized dealer.
Table 16 - Troubleshooting with app. (next)