Lightwave UV/Vis Diode Array Spectrophotometer
36
12 Servicing and Maintenance
The Lightwave has been designed to be extremely robust and require
minimal servicing.
The unit performs a series of self-diagnostic checks each time it is switched
on to check the:
•
Lamp performance
•
Wavelength calibration
•
Diode Array
The only user serviceable parts are the lamps, fan and cuvette chamber.
12.1 Replacing the Lamps
The Lightwave uses novel technology that will ensure that the lamp
replacement is a very infrequent event (the lamps should last many
thousands of hours).
The innovative method of operation requires a particular type of lamp.
The lamp alignment is also crucial to the correct operation and we
cannot guarantee performance to the published specification if lamps
other than those from WPA are used.
The part numbers for replacement lamps are as follows: -
Tungsten Lamp
S2000T
Tungsten & Deuterium Lamp
Assembly
S2000L
As for S2000L but accepting
only unit in part exchange
S2000LX
The lamps come complete with comprehensive fitting instructions.
CARE: The lamps may be hot, and fingermarks may burn to the surface
rendering the bulb useless.