biochrom Libra S50 Health & Safety Manual Download Page 10

 

 

8

 

 

advised to have the lamps changed. Deuterium and Tungsten lamps may be replaced by the 
user. Xenon lamp replacement can only be carried out by a qualified service engineer.  

 

To replace a Deuterium or Tungsten lamp ; 

o

 

Switch the instrument off, disconnect the power supply cord and allow lamps to cool. 

o

 

Locate the lamp access cover at the rear of unit and remove the two top screws and slide 
top cover back and lift off.  Never operate the unit with the lamp housing cover removed. 
 

 

 

o

 

Remove the lamp and dispose of in accordance with local regulatory practice. 

o

 

Follow the handling instructions supplied with the lamp. Do not touch the glass envelopes 
of the replacement bulbs directly. 

o

 

Replace the lamp cover and screw securely. 

o

 

Attach power cord, switch on and wait at least 30 minutes for the unit to warm up. 

o

 

Perform a new instrument baseline (on all bandwidth settings if variable bandwidth unit) 
and save this as the permanent baseline (see the online help or user manual). 

o

 

Reset the lamp hours after replacing the lamp (see the online help or user manual). 

 

TROUBLESHOOTING 

PROBLEM 

POSSIBLE CAUSE 

 

Start up calibration fails 

 

Check sample (and reference) beam is clear before switch on 
Possible optical failure, contact service support 

Negative absorbance reading 

 

Check that sample and reference cells have not been swapped 
Check sufficient height of sample solution (beam height is 
nominally 15 mm 
above the cell base for normal measurements and it is 
recommended that it be filled to 20 mm above the base) 

Unexpected results 

Check for bubbles in solution 

Absorbance values higher than 
expected 

 

Check for use of incorrect cell type 
Check cells are free from finger prints 
Check cells for contamination 
Check cell orientation 
Check reference used 
Possible optical alignment problem, contact service support 

Absorbance value lower than 
expected 

 

Check sufficient height of sample solution 
Check sample and reference are not the same 
Check sample compartment lid is properly closed 
Possible stray light issue, contact service support 

Instrument will not reference 

 

Check orientation of cuvette 
Check for use of incorrect cell type 

Poor reproducibility with 
nucleic acid analysis 

Check for particles, try using background correction 
Try increase in concentration 

Summary of Contents for Libra S50

Page 1: ...Biochrom Libra S50 S60 S70 S80 UV Visible Spectrophotometers Health Safety Manual...

Page 2: ...ardous Substances in Electrical and Electronic Equipment ROHS 2006 42 EC Machinery directive 1999 5 EC Radio and Telecommunications Terminal Equipment Directive instruments fitted with Bluetooth acces...

Page 3: ...Installation 4 Instrument Connections 5 Equipment Operation 6 Controls and Indicators 6 Intended Users 7 Instrument Preparation 7 Post Run Procedures 7 Performance Validation 7 User Maintenance 7 Tro...

Page 4: ...ard Electrical equipment for measurement control and laboratory use Classified as basic immunity criterion A and Class B for conducted and radiated emissions EN 12100 1 2 2003 A1 2009 Safety of machin...

Page 5: ...sure that the user of the equipment is provided with a safe working environment Any chemicals used in Analyses should be used stored and disposed of in accordance with manufacturer s guidelines and lo...

Page 6: ...hen it should be allowed to come to thermal equilibrium for 2 to 3 hours before operation so that start up calibration is not compromised by condensation The equipment must be connected to the local s...

Page 7: ...5 Instrument Connections...

Page 8: ...m up and stabilise for at least 30 minutes If applicable connect the unit to a PC using a USB cable and refer to the online help and user manual Select the appropriate application or method Where rele...

Page 9: ...nd solvents may be classified as hazardous or bio hazardous waste The disposal of such substances must be carried out in accordance with local regulatory practice Performance Validation Good laborator...

Page 10: ...the online help or user manual TROUBLESHOOTING PROBLEM POSSIBLE CAUSE Start up calibration fails Check sample and reference beam is clear before switch on Possible optical failure contact service supp...

Page 11: ...eb site or your local distributor Return the unit together with a completed declaration of contamination status form Available from the web site or your local distributor Please note that instrumentat...

Page 12: ...10 41 56 1872 issue 01...

Reviews: