background image

IN 45 100 010 • AA • 3 / 12 

Instructions 

Purification  can  be  carried  out  at  room  temperature  or  at  temperatures  down  to  4°C.  Operation  at  a  low 
temperature may require a  reduced  flow rate due to  the  increased  viscosity  of the buffer. All steps can be 
carried out with a syringe, a peristaltic pump or a chromatography system. If the chromatography system has 
a pressure limit functionality, set the maximum pressure over the column (resin bed) to 3 bar (remember to 
take the system fluidics contribution to the pressure into account). 
 
1. Prepare the sample 
After cell disruption or extraction, clarify the sample by centrifugation at 10 000 - 20 000 × 

g

 for 15 - 30 minutes. 

It is generally recommended to pass the sample through a 0.22 - 0.45 µm filter (e.g. a syringe filter) to avoid 
inadvertently applying any remaining particles onto the column. If the sample contains only small amounts of 
particles, centrifugation may be omitted and it is enough only to carry out filtration. Application of a sample 
that has not been properly clarified may reduce the performance and lifetime of the column. The sample should 
be applied under conditions similar to those of the binding buffer.  
 
2. Connect the column 
Cut off or twist off the end at the outlet of the column, see Figure 2. Note: It is of high importance to cut off the 
tip at the very end of the cone, preferable using a scalpel. Incorrect removal of the end piece will affect the 
performance of the column. 
 
Connect  the  column  to  your  equipment  using  the  recommended  connectors  shown  in  Table  1.  Fill  the 
equipment with deionized water or buffer and make drop-to-drop connection with the column to avoid getting 
air into the column. Carry out all steps, except for sample application, at 1 ml/min (BabyBio 1 ml column) or 5 
ml/min (BabyBio 5 ml column). 
 
 
 

 

Figure 2. Removal of the cut-off end at the column outlet should be done by cutting or by twisting (A) not bending (B). 

 

Table 1. Recommended connectors for coupling BabyBio columns to the equipment of choice. 

Equipment 

Accessories for connection 

Syringe 

Female luer or male coned 10-32 threads 

Chromatography system 

Fingertight connectors (coned 10-32 threads) for 1/16” o.d. tubing 

 
 

 

Summary of Contents for BabyBio DEAE

Page 1: ...r basic proteins i e proteins with a high isoelectric point pI while BabyBio Q and BabyBio DEAE columns are suitable for purification of acidic proteins i e proteins with low pI 1 Choose a suitable pH and buffer for the binding of the target protein One pH unit below pI BabyBio S columns or above pI BabyBio Q and BabyBio DEAE columns is a good starting point 2 Connect the column to the chromatogra...

Page 2: ...therefore interact with an ion exchange resin also at the pI The likelihood of binding to either the cation or the anion exchange resin will increase when moving away from the pI Ion exchange chromatography begins with equilibration of the column to establish the desired pH and charging the resin with suitable counter ions to the charged ligands on the resin e g the negative sulfonate groups can i...

Page 3: ...f a sample that has not been properly clarified may reduce the performance and lifetime of the column The sample should be applied under conditions similar to those of the binding buffer 2 Connect the column Cut off or twist off the end at the outlet of the column see Figure 2 Note It is of high importance to cut off the tip at the very end of the cone preferable using a scalpel Incorrect removal ...

Page 4: ...ate may reduce the yield Applied samples should have a pH that gives the target protein a charge that is opposite the charge of the column resin The pH together with the ionic strength in the sample solution might need adjustment for optimal binding 6 Wash After sample application remove unbound impurities by washing the column with 20 30 CV washing buffer or until desired A280 nm absorbance of th...

Page 5: ...m pressure limit onto the first column If possible the maximum pressure of the chromatography system should be set according to Table 2 Remember always to take the system fluidics contribution to the pressure into account Table 2 Recommended maximum pressure settings for BabyBio columns connected in series Notice that the maximum pressure over each column is always 3 bar No of columns in series Ma...

Page 6: ...fication using BabyBio S BabyBio Q and BabyBio DEAE Other buffers can possible be used Buffer Product Buffer composition Binding buffer BabyBio S 50 mM Na phosphate pH 7 0 Binding buffer BabyBio Q 50 mM Tris HCl pH 7 4 Binding buffer BabyBio DEAE 50 mM Tris HCl pH 7 4 Elution buffer BabyBio S 50 mM Na phosphate 1 M NaCl pH 7 0 Elution buffer BabyBio Q 50 mM Tris HCl 1 M NaCl pH 7 4 Elution buffer ...

Page 7: ... jam and thus reduced diffusion rate into the pores A slightly elevated salt concentration reduces but does not eliminate the interactions with the resin by creating a dynamic adsorption desorption equilibrium that allows further diffusion into the resin thus increasing the binding capacity Tuning the flow rate Flow rate is another factor that can be optimized to improve the binding capacity durin...

Page 8: ...endent on purity and recovery requirements as well as properties of the target protein and the sample Using a gradient elution gives increased purity than step elution but step elution may be necessary to obtain the highest possible concentration of the target protein In order to optimize the salt concentration for step elution an initial gradient test run can be carried out to identify a suitable...

Page 9: ...as ion exchange chromatography This can be carried out quickly and easily using BabyBio Dsalt 1 ml or 5 ml columns see Related products BabyBio Dsalt columns are also a useful alternative to dialysis for larger sample volumes or when samples need to be processed rapidly to avoid degradation Additional purification Ion exchange chromatography is a powerful single protein purification step or combin...

Page 10: ...sin using 1 M NaOH applied by a low reversed flow for 2 hours or overnight is often sufficient CIP of the column can be carried out as followed 1 Wash the column with 5 CV deionized water 2 Apply 3 10 CV of 0 5 1 M NaOH for 15 30 minutes Note The contact time is the important factor treatment with NaOH overnight can be necessary if severely fouled 3 Wash the column with 5 10 CV deionized water or ...

Page 11: ... ml 1 ml 5 ml 1 ml 5 ml Column dimension 7 x 28 mm 1 ml 13 x 38 mm 5 ml 7 x 28 mm 1 ml 13 x 38 mm 5 ml 7 x 28 mm 1 ml 13 x 38 mm 5 ml Recommended flow rate BabyBio 1 ml BabyBio 5 ml 1 ml min 150 cm h 5 ml min 225 cm h 1 ml min 150 cm h 5 ml min 225 cm h 1 ml min 150 cm h 5 ml min 225 cm h Maximum flow rate BabyBio 1 ml BabyBio 5 ml 5 ml min 780 cm h 20 ml min 900 cm h 5 ml min 780 cm h 20 ml min 9...

Page 12: ... ml 1 x 1 ml 2 x 1 ml 5 x 1 ml 10 x 1 ml 45 200 101 45 200 102 45 200 103 45 200 104 BabyBio S 5 ml 1 x 5 ml 2 x 5 ml 5 x 5 ml 10 x 5 ml 45 200 105 45 200 106 45 200 107 45 200 108 BabyBio Q 1 ml 1 x 1 ml 2 x 1 ml 5 x 1 ml 10 x 1 ml 45 100 101 45 100 102 45 100 103 45 100 104 BabyBio Q 5 ml 1 x 5 ml 2 x 5 ml 5 x 5 ml 10 x 5 ml 45 100 105 45 100 106 45 100 107 45 100 108 BabyBio DEAE 1 ml 1 x 1 ml ...

Reviews: