CI-900 Instruction Manual 8/10/2015
1554 NE 3
rd
Ave, Camas, WA 98607, USA Phone: (360) 833-8835
sales@
cid-inc
.com
www.cid-inc.com
63
CI-900 Firmware Update Procedure
1.
After installing the driver, open the
CI-900 Firmware
Upgrade
application: “Transfer Code to CI-203 or CI-
900.”
2.
If device is on, press power button to power off the
device.
Hold the down arrow button, then press power
button.
The device's internal green LED near USB
port/SD card should start flashing quickly. The device
must be unplugged for this to work.
3.
Select File and Open.
4.
Navigate to the
CI-900-ver-x.xx.bin.
This is the code for
the instrument.
5.
Follow the instructions in the software to hold down the
power button.
6.
Release the power button when instructed.
7.
Wait for the flash to be erased and then written. When the firmware transfer is complete,
the software will indicate reset sent, followed by “leaving open”. The CI-900 will power off.
8.
Power on the CI-900 and check the firmware version at
the top of the display. The Transfer Code software
application will indicate that no instrument is found, if
the CI-900:
is not connected to the computer
is not powered on
driver is not properly installed
If this message is seen, make sure:
the CI-900 is connected via mini-USB
Summary of Contents for CI-900