Section 4: Optional Setting GC Emulation Mode Calibration Manually
If performing a calibration injection and verifying the accuracy of the calibration, it is no longer
necessary to carry out the following steps. If questions arise as to how the calibration is working
and what the area being used for the peak counts is, they can be found here. However, it is
recommended to run the standard calibration injection protocol (previous GC emulation sections 1-
3) and have the CI-900 automatically provide the peak counts.
1. Take 3 measurements of a calibration gas standard.
2. Determine the average of the 3 peak counts.
3. Set the value for peak counts to the average and enter the ppb concentration in the
calibration menu. (Setup > Calibration > press Enter to continue > C
2
H
4
> Standard >
Injection parameters).
Injection parameters
peak counts 1103698
concentration 0001000
calibrated Volume 50.0