keys.

6.3 Flow Rate Calibration

4.

To change the direction of flow, stop the pump by press-

i ng the Run/Stop key. Press the Direction key to choose

the new direction. Restart by pressing the Run/Stop key.

5.

To purge the tubing, press the Purge key. The pump will
run at 100% maximum speed. When the tubing is
primed, press the Purge key again to stop the purge
function and return to the Run mode, or press the
Run/Stop key to turn the pump off.

An uncalibrated Model EP- I Econo Pump will display pump speed as per-

cent of maximum rpm. The user may also choose to select one of three tub-
ing sizes for which calibration has been programmed at the factory. The
Model EP-1 Econo Pump also features a user calibration mode, which is
used when improved flow rate accuracy is desired, or when using non-stan-
dard size tubing. The pump must be calibrated using one of the methods

described above before a fraction collection program may be entered.

Flow Rate Calibration with Pre-Selected Tubing Sizes (0.8, 1.6,

and 3.2 mm ID)

6.2 Verification of Econo System Software Version

The Model ES-1 Econo System Controller and Model EP-1 Econo Pump

should already be set up.

The Model EP-1 Econo Pump must have software version 2.01 or
higher to control the Model EV-1 Econo Buffer Selector.

If you do not have a Model EV-1 Econo Buffer Selector, continue with

Section 6.3. Otherwise check the software version of your system:

Simultaneously press and hold the Direction Key and the "down"
Arrow key on the front panel of the Model EP-1 Econo Pump. The
four digit LED display on the front panel of the Pump should dis-

play: "v2.01" or higher.

If your pump displays a number lower than 2.01 , and you will be
using the Model EV-1 Econo Buffer Selector, your pump software
should be upgraded. Contact your local Bio-Rad representative for

details on how to upgrade software.

If your pump currently features software version 2.01 or higher ,
insure that all connections between the Model EP-1 Econo Pump,
Model ES-1 Econo System Controller, (and Model EV-1 Econo
Buffer Selector, if used) have been completed and the system power
is on.

1 2

User Calibration Procedure

1.

With the pump in Stop mode, install tubing as described
in Section 5.2.

2.

Press the key in the lower left-front of the pump panel
corresponding to the pump calibration feature, and con-
tinue pressing until the indicator light corresponding to

either 0.8, l.6, or 3.2 mm ID tubing is lit.

3.

The LED display reads ml/min for any of these pre-pro-
grammed tubing settings. If improved flow rate accuracy
is desired, see User Calibration Procedure below.

4.

The pump is now automatically calibrated for the specif-
ic tubing size. Press the Run/Stop key to start flow.
Increase or decrease the pump speed with the Arrow
keys.

1.

Install tubing as described in Section 5.2. Connect pump
inlet to a container of water or buffer. Connect a length

of tubing to the pump outlet.

2.

Run pump to purge air from tubing, then stop pump.
Place the end of the outlet tubing in an empty graduated
cylinder.

3.

Press Calibration key until "CAL" is displayed. Press the
Run/Stop key. Pump will run for 5 minutes at 25% of
maximum speed, and the display will count down time.

13

Summary of Contents for EP-1 Econo Pump

Page 1: ...Model EP 1 Econo Pump Instruction Manual Catalog Numbers 731 8140 731 8142 For Technical Service Call Your Local Bio Rad Office or in the U S Call 1 800 4BIORAD 1 800 424 6723...

Page 2: ...8 5 l Tubing Selection 8 5 2 Installation of Precut Bio Rad Tubing 9 5 3 Installation of Uncut Tubing 11 Section 6 Operation 11 6 l Start Up Procedure I1 6 2 Verification of Pump Software Version 12...

Page 3: ...Section 1 Safety Disconnect supply before servicing No user serviceable parts inside refer servicing to Bio Rad service personnel 1...

Page 4: ...Econo Pump will control the gradient run time gradient shape and peak collection capabilities of the Econo System When coupled with the Model ES 1 Econo System Controller and Model EV 1 Econo Buffer...

Page 5: ...ration Rotate the voltage selector through the window in the fuse drawer until the proper voltage shows 4 Reinsert the fuse drawer into the power entry module with the locking tab to the left The volt...

Page 6: ...rter valve to collect or waste positions and total run volume The total run vol ume V may be used to program the pump to automatically shut off after a given volume has been delivered 4 2 Rear Panel F...

Page 7: ...hree and will provide the most consistent flow rate over time Table 5 l shows approximate flow rate ranges with different tubing sizes Flow rates above 20 ml min may be obtained by plumbing two channe...

Page 8: ...onto the pump head as described in Section 5 2 Warning If using tubing other than the type supplied by Bio Rad make sure the wall thickness is not greater than 1 0 mm Using tubing with a greater wall...

Page 9: ...nt panel of the Model EP 1 Econo Pump The four digit LED display on the front panel of the Pump should dis play v2 01 or higher If your pump displays a number lower than 2 01 and you will be using the...

Page 10: ...g prompting entry of fraction size Enter the fraction size in 0 l ml increments using the Arrow keys After frac tion size is selected press the Fraction Collector key once more to confirm the fraction...

Page 11: ...display will show time elapsed The pump will stop automatically when the total run volume has been delivered 1 6 7 To abort the fraction collection scheme at any time dur ing the program press the Run...

Page 12: ...o continue a run in FAIL progress press program run key Otherwise press the run stop key Pump will not run No power Verify that powercord is plugged no LEDs lit in and the power switch is turned on Fu...

Page 13: ...than 15 milliamperes of current The following table describes the two rear panel mini DIN connector pinouts Table B Rear Panel Pinouts 21 Technical Specifications Number of channels 2 Flow rate range...

Page 14: ...cks and tub ing connectors Cables 731 8261 Econo System Cable 1 8 pin mini DIN to DB 9 connector For connection of Model 2110 Econo Fraction Collector to the Econo Pump Econo UV Monitor or Econo Syste...

Reviews: