Replacing the fuses
If the instrument doesn’t start, check that the electric cable has been connected properly, there is current and the
two fuses under the instrument’s socket aren’t burnt. If necessary, change them with fuses of identic value.
Fuses holder socket (N. 2 delayed fuses of 3,15 Ampere - T3.15AH250V).
Instructions for the replacement: Switch off the appliance and unplug the cable from the socket. Make a light
pressure on the little carter covering the fuses (if necessary using a little screwdriver), change them, close the carter
and check the instrument’s ignition.
Replacing the lamp
Switch off the instrument and unplug the cable from the socket.
Remove the panel at the bottom of the instrument unscrewing the 8 fixing screws and lift the chassis rotating it on
the left. Rotate the lamp of 90° and
lift it. Change it with an equal model and rotate it of 90°. Close the instrument
again.
Replacing LAE thermostat
Switch off the instrument and unplug the power supply cable from the socket, open the instrument. The thermostat is
fixed with clips or blocking flasks. The connections are on the label positioned on the upper side of the thermostat.
Open the fuses box
using a screwdriver
Fuse box opened