After  the  Microblaster  installation  has  been  finished,  correctly  follow  the  use 
instructions and the other information about precautions and safety, described below.

7. USER INSTRUCTIONS

1

-Remove the switch button clamp (04). 

_______

When removing the switch button clamp (04), try using it immediately

 

_______

afterwards  and  always  after  using  Microblaster,  return  the  respective

 

_______

clamp  to  the  button  (04)  thus  keeping  it  always  pressed.  The  act  of

 

_______

keeping  the  button  clamp  released  for  long  periods  might  generate

 

_______

adherence of the internal driving tube walls, which might make the oxide

 

_______

passage difficult.

2

-Microblaster  Plus  model  is  provided  with  two  removable  and  autoclavable 

nozzles (02 and 11) offering two options for the ejecting needle angling, one at 90º 
(indicated for intraoral works on posterior teeth) and another at 138º (indicated for 
laboratory works and intraoral works on anterior teeth).   According to the kind of 
work to be carried out, fix one of the nozzles on the Microblaster body by means of 
the fixing nut (12).

_______

During the nozzles removal or fixation (02 or 11) on Microblaster Plus, 

_______

firmly hold the nozzle with the fingers of one hand without turning it and 

_______

then turn the connecting nut with the fingers of the other hand (12), thus 

_______

fixing or releasing the chosen nozzle by means of the nut thread.

_______

Never  use  any  kind  of  wrench  or  tool  to  fix  or  release  the  nozzle  on 

_______

Microblaster Plus.

Note: 

Nozzle (02) used in the Microblaster models allows rotation up to 180º, 

facilitating the performance of very different kinds of works.

3

- Fill the tank (06) in to maximum ¾ of its capacity with aluminum oxide and fix it 

on the tank for abrasive lid (5) located at the end of the Microblaster body.

_______

Use only abrasives of the type aluminum oxide and with granulations in 

_______

the range up to 100 microns.

Note: 

Bio-Art commercializes the perfect aluminum oxide for consumption and 

refill to be used with Microblaster.

5

Summary of Contents for Microblaster

Page 1: ...MicroBlaster i n t e l l i g e n t s o l u t i o n s USER MANUAL...

Page 2: ...lation instructions 7 Use instructions 8 Maintenance and cleaning 9 Safety precautions 10 Warnings 11 Special storage conditions 12 Problems and solutions 13 Symbols 14 Term of warranty and technical...

Page 3: ...Microblaster Standard and Lab Models One xed nozzle mounted at a 138 angle Microblaster Plus Model Two removable and autoclavable nozzles mounted at 90 and 138 angles Attention Before operating the eq...

Page 4: ...138 Yes Yes No No No Yes No No Yes Yes Yes Yes Yes No Yes No Yes No Removable nozzles with ejecting needles at 90 and 138 Autoclavable nozzles Use of aluminum oxide with granulation up to 100 microns...

Page 5: ...t and the user s safety and for better performance during the blasting works make sure the compressed air pressure is kept stable at 60 to 80 lbf inch2 4 5 5 Kgf cm Bio Art supplies the connections 07...

Page 6: ...d intraoral works on anterior teeth According to the kind of work to be carried out x one of the nozzles on the Microblaster body by means of the xing nut 12 _______During the nozzles removal or xatio...

Page 7: ...constant movements throughout the whole surface to be blasted Some recommendations for the aluminum oxide granulation indicated for some application examples are presented below It is upon the profes...

Page 8: ...croblaster cleaning and disinfection use only a cloth dampened with water or alcohol 70 _______Bio Art Microblaster shall never be sprayed with liquids be submerged _______or sterilized in an oven or...

Page 9: ...goggles gloves and breathing mask when operating the equipment In addition to these EPI s and for cases when the work offers such conditions we recommend the use of the Bio Art Bio Chamber suction ch...

Page 10: ...s user manual and monitor the correct pressure 60 to 80 lbf inch 4 5 5 Kgf cm Fill the tank 06 in with aluminum oxide at maximum of 3 4 of its capacity The compressed air network shall be dry Check th...

Page 11: ...s all and any manufacturing defect being provided through repair of the equipment and subjected to the following requirements That the product has been used correctly in accordance with the instructio...

Page 12: ...637 034 447 113 _______www bioart com br 15 AUTHORIZED TECHNICAL ASSISTANCE 16 EUROPEAN AUTHORIZED REPRESENTATIVE For your safety technical assistance on this product shall only be provided by authori...

Page 13: ......

Page 14: ...Rev CMAN0274 f Fev 2016 i n t e l l i g e n t s o l u t i o n s Rua Teot nio Vilela 120 Jd Tangar CEP 13568 000 S o Carlos SP Brasil Tel 55 16 3371 6502 Fax 55 16 3372 5953 www bioart com br...

Reviews: