CNCR-110 • Two-wire 4-20
15
Failure
Function check
Code
Text message
Cause
Rectification
F013
no measured value avail-
able
No measured value in the boot up phase
or during operation
Check or correct installation
and/or parameter settings
Clean the antenna system
F017
Adjustment span too small
Adjustment not within specification
Change adjustment according to the
limit values (difference between min.
and max. ≥ 10 mm)
F025
Error in the linearization
table
Linearization values are not continuously
rising, for example illogical value pairs
Check linearization table
Delete table/Create new
F036
No operable software
Checksum error if software update failed
or aborted
Repeat software update
Send instrument in for repair
F040
Error in the electronics
Limit value exceeded in signal processing
Hardware error
Restart instrument
Send instrument in for repair
F080
General software error
General software error
Restart instrument
F105
Determine
measured value
The instrument is still in the boot up
phase, the measured value could not
yet be determined
Wait for the end of the boot up phase
Duration up to 3 minutes depending on
the measurement environment and
parameter settings
F260
Error in the calibration
Checksum error in the calibration values
Error in the EEPROM
Send instrument in for repair
F261
Error in the instrument
settings
Error during setup
False signal suppression faulty error
when carrying out a reset
Repeat setup
Reset instrument
F265
Measurement function
disturbed
Program sequence of the
measuring function disturbed
Device restarts automatically
Code
Text message
Cause
Rectification
C700
Simulation active
A simulation is active
Finish simulation
Wait for the automatic end after 60 mins.