Redline Operating Manual RE / RF / RI 12/2010
page 35/46
Press the arrow key until you call up the upper adjustment value. If you want to enter a reference value
for the upper adjustment value, press the functional button. The display automatically reverts to the main
menu.
Adjustment menu
(lower adjustment
value)
Adjustment menu
(upper adjustment
value)
Main menu
(e.g. Standby)
Call up again the adjustment menu. “C.ADJ” is displayed. Enter the value, which has been measured with
the reference measuring device at the upper adjustment value, using the arrow keys. The controller takes
over the value and shows again the adjustment menu (selection menu).
Example for upper adjustment value: controller shows 120 °C, reference measuring device shows 122
°C.
Adjustment menu
(reference value entry)
Adjustment menu
(reference value entry)
Adjustment menu
(selection menu)
5 x
or
CALIBRATION
ADJUST
CALIBRATION
ADJUST
USER
CALIBRATION
Back to the main menu display: press the functional button or wait for 1 minute.
The temperature controller of your redLINE chamber has now been adjusted.
Resetting the adjustment:
Adjustment menu
(selection menu)
Adjustment menu
(Reset)
Adjustment menu
(selection menu)
3 x
1 sec.
USER CALIBRATION
USER CALIBRATION