KMF (E5.2) 06/2014
page 37/99
6.3 Menu settings in the “User-settings” menu
User-settings
Instrument data
Contrast
Displ. Power down
Contin. operation
User Level
Safety control. Act + 39.8 °C
Safety control. Set. + 42.0 °C
29
Instrument data
Instrument Name
Enter an individual name of the constant climate chamber.
Address
Enter a controller address (1 to 30) for operation with the communication
software APT-COM™.
All other entries are relevant only for service purposes.
Contrast
(no function)
Displ. power down
Switch off event
Do not change the entry “Wait. Period”.
Waiting period
You can enter a delay time after which the display, following manual activa-
tion, will automatically be turned off. This happens when the moment is out-
side the operation time defined in menu ”Contin. operation”.
Contin. operation
Enter an operation time to determine the period of display activity. Outside
the defined time, the display is automatically turned off. Pressing down any
key will reactivate the display. After the time set in menu “Displ. power down“,
the display will turn off again when the actual time is not within the operation
time fixed in menu “Cont. operation“.
User Level
Toggle here to the display menu “User Level” (chap. 6.4) by entering a pass-
word. Factory default setting for this password is +00001. You can change
the password (“user code”) in the menu “User Level”.
Safety control.Set
The setting of the tolerance limit of the safety controller (chap. 10.2) is dis-
played. You cannot change it in this view.
Safety control.Act
The measuring result of the safety controller (chap. 10.2) is displayed. The
safety controller compares the value measured by a second independent
temperature sensor to the entered tolerance limit.