FED (E2) 12/2014
page 17/55
4.
Installation
4.1 Electrical connection
The drying and heating ovens FED are supplied ready for connection. The socket must also provide a
protective conductor.
•
FED 53, FED 115, FED 240:
Shockproof plug, power supply voltage 230 V (1N~) +/- 10 %, 50/60 Hz
Fixed power connection cable of 1800 mm in length
•
FED 400, FED 720:
CEE plug 5 poles, power supply voltage 400 V (3N~) +/- 10 %, 50/60 Hz
Fixed power connection cable of 2700 mm in length
•
FED 53-UL, FED 115-UL:
NEMA plug 5-20P, power supply voltage 115 V (1N~) +/- 10 %, 60 Hz
Fixed power connection cable of 1800 mm in length
•
FED 240-UL, FED 400-UL, FED 720-UL:
NEMA plug L21-20P, power supply voltage 208 V (3N~) +/- 10 %, 60 Hz
Fixed power connection cable of 2700 mm in length
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the unit’s type plate (unit front behind the door, bottom left-hand, chap. 1.4).
•
When connecting, please observe the regulations specified by the local electricity supply company
and as well as the VDE directives (for Germany). We recommend the use of a residual current circuit
breaker.
•
Pollution degree (acc. to IEC 61010-1): 2
•
Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap.12.4).
To completely separate the unit from the power supply, you must disconnect the power plug.
Install the unit in a way that the power plug is easily accessible and can be easily pulled in
case of danger.
4.2 Connection to a suction plant (optional)
When directly connecting a suction plant the spatial temperature exactitude, the heating-up and the
recovering times and the maximum temperature will be negatively influenced. So no suction plant should
be directly connected to the exhaust duct.
Active suction from the oven must only be performed together with extraneous air. Perforate
the connecting piece to the suction device or place an exhaust funnel at some distance to the
exhaust duct.