C 150 (E2) 06/2011
page 41/78
Select the desired set point type.
After 30 seconds the controller reverts to Normal Display automatically.
9.2.2 Setting the safety controller set point
You can check and set the safety controller set point in operating modes HAND (chap. 7.2) or USER
(chap. 7.3).
•
With set point type “Limit” (Lit), operating function “
AL.S
” (safety controller set point as limit value) is
displayed in operating modes HAND and USER. You can set a limit value.
•
With set point type “Offset” (OFS), operating function “
Ot.S
” (safety controller set point as offset value)
is displayed in operating modes HAND and USER. You can set an offset value.
The settings of the safety controller are inactive during sterilization (chap. 15.4). They become
functional again following abortion of the sterilization and/or the restart of the unit at the main
power switch.
Normal Display
simultaneously for 3 seconds
Toggle to operating mode USER
Operating mode HAND
Operating mode USER
If operating mode HAND appears,
toggle to operating mode USER with
the arrow keys.