CB-S / CB-S-UL (E7) 06/2020
page 71/119
•
Other event messages according to existing alarms
•
The moment of switching the alarm state on and off is indicated
under “On/Off”.
17.2 Storage capacity
The storage capacity of the data recorder depends on the number of entries.
•
DL1 = 110.000 entries (equaling 76 days with a storage rate of 1 minute, setting see chap. 17.3)
•
DL2 = 27.000 entries (equaling 18 days with a fix storage rate of 1 minute)
•
Event list: 200 events
The shorter the set storage rate, the closer are the stored measuring points, the more precise, but also
shorter is the documented period.
Once the storage capacity of the data recorder is reached, overwriting of the oldest values begins.
17.3 Setting the storage rate for the “DL1” recorder data
Required access level: “Admin”.
Path:
Normal display
Settings
Data recorder
Storage interval
Press the
OK button
to enable the setting.
Function “Storage interval”.
The current setting flashes. Use the
arrow buttons
to enter the desired
storage interval. Setting range: 1 minute to 60 minutes.
Press the
OK button
to confirm the setting.
Storage interval
With the
Back button
you can go back to the “
Data recorder
” submenu and, repeatedly pressing it, to
Normal display
.
17.4 Deleting
the data recorder
When importing a configuration via USB
stick (chap. 18.2) and when loading a new firmware version by
BINDER service, the entire data memory is deleted.
BINDER service can also install the configuration by means of a setup program without deleting the data.
Regardless of this, BINDER Service
can delete the data via a setup program.
Loading a new configuration via USB-Stick leads to deleting the data recorder.
HINWEIS
Danger of information loss when loading a new configuration.
Data loss.
Backup data on the USB stick prior to loading a new configuration.
Summary of Contents for 9040-0189
Page 109: ...CB S CB S UL E7 06 2020 page 109 119 26 9 Dimensions CB S CB S UL 170 Dimensions in mm...
Page 110: ...CB S CB S UL E7 06 2020 page 110 119 CB S CB S UL 260 Dimensions in mm...
Page 112: ...CB S CB S UL E7 06 2020 page 112 119...
Page 113: ...CB S CB S UL E7 06 2020 page 113 119 28 Product registration...