Operating functions in operating mode USER:
Display Setting range
Operating function
-90 °C to -40 °C
Temperature set-point
On OFF
Release of CO
2
emergency cooling
Switch on or off CO
2
emergency cooling
-70 °C to -20 °C
Set-point entry of CO
2
emergency cooling
(only with option CO
2
emergency cooling “On”)
On
Switch on the test of CO
2
emergency cooling
(only with option CO
2
emergency cooling “On”)
The function automatically turns off after 10 seconds.
-90 °C to -40 °C
Safety controller set point (Limit value)
Limit value, i.e., maximum permitted absolute temperature value. When
exceeded, the safety controller triggers an alarm. Set the limit value by
approx. 15 K above the controller temperature set point.
0 °C to 50 °C
Setting of ambient temperature alarm
(factory setting 33 °C)
On OFF
Switch on or off ambient temperature alarm
Indication of supply voltage
(only with option Advanced voltage boost-
er)
0 to 999
Max. 3-digit numeric
password
for locking/unlocking the operating func-
tion setting in operating mode LOCK (chap. 8.4)
To permit temporary unlocking: select any password other than 0.
To permit permanent unlocking: set password to “0”.
Summary of Contents for 9020-0213
Page 6: ...UF V E2 E2 1 03 2013 page 6 112...
Page 7: ...UF V E2 E2 1 03 2013 page 7 112 Product registration...
Page 104: ...UF V E2 E2 1 03 2013 page 104 112 18 7 Dimensions UF V 500 E2 xxx Indications in mm...
Page 105: ...UF V E2 E2 1 03 2013 page 105 112 18 8 Dimensions UF V 500 E2 1 xxx Indications in mm...
Page 106: ...UF V E2 E2 1 03 2013 page 106 112 18 9 Dimensions UF V 700 E2 1 xxx Indications in mm...
Page 113: ......