You can enter
Program sections
into this program table.
Press the “PGM” button. An inquiry display appears allowing you to enter or delete individual program
sections:
ZP-Abschnitt Abs. Nr. 5
insert
delete
In this view, new program lines can be entered or deleted:
new
New lines are added below in the table
insert
New lines are added above a previously selected line
delete
Individual lines that have been selected previously are deleted
Create as many lines, i.e. program sections, as desired. As a next step, values can be entered into these
lines. It is possible to add supplementary lines later or to delete individual lines at any time.
PGM
08:43:55 15.02.02
PGM
08:43:55 15.12.05
To enter values, select the corresponding line via the
arrow keys.
Press the “ENTER” button. The
program editor
ap-
pears.
Enter the individual values of the selected program section.
Program editor Abs.Nr. 6
****.*
Setpoint 1
+100.0
FAN
Operating contacts
Time
00:45:00
Repeat Section
5
Repeat Number
10
Tol.-band min.
-1999.0
Tol.-band max.
+9999.0
Parameter set
1
-- Temperature value at the start of the program section
-- Fan speed in %
-- Operating
contact
(operation
line)
1
=
Bedew protection
-- Duration of the program section
-- No. of start section in case of repeat cycles
-- No. of repeats in case of repeat cycles
-- Temperature limits (maximum / minimum temperature)
In case of exceeding: temporary program stop.
-- Pre-selected value (Do NOT change!)
Select the parameters via the arrow keys and confirm by pressing ENTER.
Then enter the values via the arrow keys, and confirm the entry by pressing ENTER.
For a negative set-point entry, enter the numerical value first, and then the minus sign (-).
With set-point type “
Limit
”, the user shall adapt the safety controller (chap. 11.2) to the highest
temperature set-point value of the program actually used. Check the safety controller for each
temperature program and change it if necessary. Set the safety controller set-point by approx..
5 °C to 10 °C above the highest temperature set-point of the program.
Recommended setting: set-point type “
Offset
” with safety controller set-point 5 °C.
In case of the optional over-/under temperature safety device (chap. 11.3), check also the
temperature limits entered there, and adjust them if necessary.