N
o
7 - Drawing OPTIMA INT connection with DMX3 1 motor
without holder
OPTIMA
INT
ENG
ITA
42
Voltage converter
24 VAC / 32 - 24 VDC
REF 1500580
(Supplied)
+3 2Vdc
Gnd
OPTION B: MCX + MX2 (32VDC / 5A)
Schurter Filter
KFA 4301.5206
115/230VAC 50/60Hz
(Not supplied)
110 /
230VAC
Transformer
110 / 230 Vac / 24 Vac /
150 VA
(Not supplied)
1
0
VOLTAGE CONVERTER
24 VAC / 32 - 24 VDC
REF 150 0580
32 Vdc OUT: 130 W peak / 60 W cont
24 Vdc OUT: 24 W peak / 12 W cont.
32 Vdc OUT
GND
24 Vdc OUT
GND
24 Vac IN
24 Vac IN
6
5
4
3
2
1
Integrator's
menu => setup
12
4
5
6
ON
3
+5V
(Brown)
(Orange )
(Bla ck)
(Red)
(Blue)
L +
L -
Phase C
Phase B
Phase A
Gnd
GND
Gnd
+3 2Vdc
Speed ref
10K (Pedal)
MX2 / MCX-1
HOSE
MX2 / MCX
Cooling
400kPa
OPTION C: MCX + MX2 (32VDC / 5A)
+3 2Vdc
GN D
100 /
240VAC
Power supply PMP90-16-k24
REF 1500666
OPTION A
OK
REV
SETUP
SAVE
ESC
OPTIMA
INT
Bien Air
Dental
DMX3
Without holder
Motors MX2 / MCX
SWISS MADE
MX2 motor REF 1600677
MX2 hose REF 1600700
MCX motor REF 1600751
MCX hose REF 1600756
GND
+5 V
Speed ref
0-300 kPa
DMX3 REF 1501468