T E C A L  1 4 0 S   O P E R A T O R ’ S   M A N U A L

Specification

The figures stated in the specification are as measured 0-50 mm from the base of the
well of the insert, using an SPRT, at the time of calibration.

Measuring zone

0 to 50 mm

0" to 2"

from base of well

from base of well

Depth of well in insert

114.3 mm

4.5"

Temperature range

-40°C to 140°C  *

-40°F  to 284°F *

Temperature accuracy in measuring zone

±0.3°C

±0.54°F

Temperature uniformity in measuring zone

±0.2°C at 100°C

0.36°F at 182°F

Temperature stability after 10 minutes

±0.05°C

±0.09°F

Display resolution

0.1°C

0.1°F

Heat up rate

5 minutes

20°C to 100°C

68°F to 182°F

Cool down rate

9 minutes

100°C to 0°C

182°F to 32°F

Comms Port 9way D type

Full bi-directional RS232

* The lowest temperatures can only be achieved when the Tecal 140 is used with a chilled
water supply which has a maximum pressure of 6 metres of water.   However, the chilled
water supply must not be switched on when the set temperature is above 40°C as this can
lead to damage to the block.

Dimensions

Height

272mm

10.7"

Width

185mm

7.28"

Depth

439mm

17.28"

Wells in the block:

Central Well for inserts, see the list of inserts at the back of this book

Depth

130mm

5.1"

Diameter

38.2mm

1.5"

Working conditions

The Tecal units are designed to work safely under the following conditions:

Ambient temperature range 5°C to 40°C
Humidity

Up to 95% relative humidity, non-

condensing

Note:

 The control specifications quoted are for an ambient temperature

range of 10°C to 30°C.  The specification may deteriorate outside this range
but the unit will still work safely.

Radio frequency interference tested and passed to EN50081-1.
Immunity Tested and passed to EN50082-1

Packing

When you receive your unit, make sure you keep the original packing in
case you ever need to return it for service or repair. When returning a unit,
remember to remove the insert from the temperature controlled block.

The unit must be transported in the original packing to avoid damage. The
packing comprises; the unit in the soft carrying case packed with foam into
an outer cardboard box. Techne accepts no responsibility for damage
incurred unless the unit is correctly packed and transported in this way.

Summary of Contents for TECHNE Tecal 140S

Page 1: ...Issue 4 Date of issue 3rd May 2005 Techne is a trademark Techne 2003 Tecal 140S Calibrator Operator s Manual...

Page 2: ......

Page 3: ...cations 8 Warning 8 Specification 9 Working Conditions 9 Packing 9 WHEN YOU SWITCH ON 10 Front Panel controls 10 Setting the Units for Temperature 12 Setting the Operating Temperature 12 Water Cooling...

Page 4: ...PAGE 2 T E C A L 1 4 0 S O P E R A T O R S M A N U A L...

Page 5: ...PAGE 3 T E C A L 1 4 0 S O P E R A T O R S M A N U A L NOTES...

Page 6: ...that the air inlet vents on the underside are free from obstruction 5 Note that the following symbols may be next to the indicator lamps on the front panel of the units and have the following meanings...

Page 7: ...assurant que les trous d a ration situ s sous l appareil ne sont pas obstru s 5 Les symboles ci dessous situ s c t des t moins lumineux sur la face avant de l appareil ont la signification suivante t...

Page 8: ...t bevor Sie dieses Ger t benutzen Warnung HOHETEMPERATURENSINDGEF HRLICH siek nnen dem Bediener ernsthafte Verletzungen zuf gen und brennbare Materialien k nnen sich leicht entz nden Techne hat bei de...

Page 9: ...to a las luces indicadoras en el panel frontal del equipo tienen los siguientes significados Indicador de potencia Indicador del calor Indicador de sobretemperatura 6 Los s mbolos que se encuentran en...

Page 10: ...tailed towards the back of this manual Maintaining a set temperature The required temperature is set on the calibrator and the operation of the probe is checked A NAMAS certificate can be provided con...

Page 11: ...to damage to the block Dimensions Height 272mm 10 7 Width 185mm 7 28 Depth 439mm 17 28 Wells in the block Central Well for inserts see the list of inserts at the back of this book Depth 130mm 5 1 Diam...

Page 12: ...this for 1 second then the actual temperature of the block will be indicated The Front Panel Controls The front panel controls consist of three buttons for controlling the display a five digit LED di...

Page 13: ...t indicator shows when there is power to the block When the set temperature is different to the block temperature this indicator will light If the light is on continuously the block is getting constan...

Page 14: ...ended temperature range to 40 C can be achieved on the Tecal 140 unit with a 4 C chilled water supply The supply must have no more than 6 metres of water pressure head However the chilled water supply...

Page 15: ...owered up otherwise data integrity cannot be guaranteed Once the cable is fitted it does not matter which unit is powered up first RS232 9 way PC serial port ON OFF switch Fuses Power inlet The follow...

Page 16: ...display If HELP is displayed there is a problem with the internal sensing device The calibrator should be returned to your supplier for repair Cleaning your Tecal unit Before cleaning your unit disco...

Page 17: ...cable The following parts may be purchased if replacements or alternatives are required 6101058 Insert extractor FCAB10UK Mains cable 230V UK units FCABLEEU Mains cable 230V units FCABLEUS Mains cable...

Page 18: ...ation Aluminium Inserts Immersion depth 114 3 mm or 4 5 NUMBER PROBE DIAMETER METRIC FINSALA 5 off 6mm FINSALB 1 off each 10mm 8mm 6mm 4 5mm 3mm FINSALC 2 off 6mm 2 off 10mm FINSALD 2 off 6mm 2 off 12...

Page 19: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Reviews: