1
Water Baths
sWB1, sWB1D, sWB2, sWB2D, sWB3,
sWB3D
Introduction
thank you for purchasing this piece of
stuart equipment. to get the best
performance from the equipment, and for
your own safety, please read these
instructions carefully before use.
Before discarding the packaging check that
all parts are present and correct.
These Stuart water baths are designed to
operate under the following conditions:-
❖
For indoor use only
❖
Use in a well ventilated area
❖
Ambient temperature range +5°C to +40°C
❖
Altitude to 2000m
❖
Relative humidity not exceeding 80%
❖
Mains supply fluctuation not exceeding 10%
❖
Overvoltage category II IEC60364-4-443
❖
Pollution degree 2
If the equipment is not used in the manner
described in this manual the protection
provided by the equipment may be
impaired.
electrical Installation
thIs INstrUMeNt MUst Be eartheD
Before connection please ensure that the
line supply corresponds to that shown on
the rating plate. these models require a
supply rated at 220 - 240V ac, 50hz, single
phase.
All models are supplied with two mains leads
fitted with IEC plugs for connection to the
instrument. One has a UK 3 pin plug and the
other has a 2 pin "Shuko" plug for connection
to the mains. Choose the lead appropriate for
your electrical installation and discard the other.
Should neither be suitable, take the lead with
the UK plug and replace the plug with a suitable
alternative. This involves cutting off the moulded
plug, preparing the cable and connecting to the
re-wireable plug in accordance with its
instructions.
Model
Power
SWB1/SWB1D
750W
SWB2/SWB2D
1kW
SWB3/SWB3D
1.5kW
CaUtION - Fuse fitted in both live and
neutral lines.
The wires in the mains cable are coloured as
follows:
BROWN
- LIVE
BLUE
- NEUTRAL
GREEN/YELLOW - EARTH
All models are supplied with a mains lead fitted
with an IEC plug and IEC socket in the rear of
the instrument. This lead should be connected to
the instrument BEFORE connection to the mains
supply.
Should the mains lead need replacement, a cable
of 1mm
2
of harmonised code H05W-F connected
to an IEC 320 plug should be used.
IF IN DOUBt CONsULt a QUaLIFIeD
eLeCtrICIaN