background image

17

16

Special cases

If the customer has a spectacle frame from Coselgi or BHM-Tech, we offer special adapters 
for these (Figs. 30 and 31). When using the 

pan

 with these adapters, an overall length as 

with existing BHM-Tech and Coselgi hearing temples is attained. This means no subse-
quent preparation of the existing glasses is necessary.   

Connection for programming 

Note: No battery needs to be inserted for programming.

Remove the programming socket cover and plug the connector of the programming cable 
into the programming socket (Fig. 32).

Note: 

Ensure that the 

red marker

 on the programming cable connector is on the same 

side as the

 marking on the housing 

(no connection is established otherwise)

.

Fitting software

To guarantee problem-free function, only use BHM-Tech software approved for this pro-
duct. Please refer to the „BHMFitpan“ software description for details on the programming 
process. Please contact your vendor or BHM-Tech Produktionsgesellschaft mbH if you re-
quire clarification.

After the programming process, remove the programming cable and put on the programming 
socket cover. If the hearing system is not used for a longer period, keep it in the original 
packaging without batteries in a safe place not accessible to anyone unauthorised.

Fig. 30 
COSELGI adapter (left & right)

Fig. 31
BHM-Tech adapter (usable left
& right)

Fig. 32

Summary of Contents for Pan Series

Page 1: ...Fitting Manual for pan...

Page 2: ...t hearing aid acousticians doctors and show proper fitting and adjustment of the pan hearing system Purpose Hearing systems are developed to improve hearing The key functions of a hearing sys tem are...

Page 3: ...adapter at marking B Fig 7 Determining the temple length To determine the correct temple length use the fitting template supplied Fig 1 Ensure the wearing position of the spectacles is correct Use a p...

Page 4: ...is an adapter ring with different adapters suitable for different wire cores Fig 10 Select from them the adapter suitable for your temple Fig 9 Clean the wire core and remove any grease Fig 10 C Adap...

Page 5: ...ng of the hearing system by using hot air only on the area specified in Fig 13 using a frame warmer The hearing system may only be heated uniformly in the area stipulated Please do not use any tools s...

Page 6: ...al is read to determine the receiver tube length Fig 18 Selection and mounting of the earpiece Use the condition of the auditory canal to select the earpiece suitable for your customer and fit it onto...

Page 7: ...e by pressing together the mounting tool Insert the High Power receiver into the open earpiece Carefully remove the mounting tool Ensure that the earpiece is fully on the mounting tool Removing the HF...

Page 8: ...ng the frame warmer Fig 28 Allow the receiver to cool for at least one minute and remove the spiral spring and protective sleeve Fig 29 Mounting and dismounting the receiver See Figs 23 to 25 for fitt...

Page 9: ...r on the programming cable connector is on the same side as the marking on the housing no connection is established otherwise Fitting software To guarantee problem free function only use BHM Tech soft...

Page 10: ...18 Notes...

Page 11: ...the content BHM Tech Design RABOLD UND CO Photos BHM Erwin Muik BHM Tech Produktionsgesellschaft mbH Grafenschachen 242 7423 Grafenschachen Austria www bhm tech at According to EU guidelines 93 42 EE...

Reviews: