SPECIFICATIONS
Accuracy is specified for a period of one year after calibration and at 18°C to 28°C, with relative humidity <
75%
Accuracy specifications take the form of:
± ([% of Reading] + [number of Least Significant Digits])
DC Voltage
Range
Resolution
Accuracy
± (0.8% + 5)
6 V
0.001 V
60 V
0.01 V
600 V
0.1 V
1000 V
1 V
± (1.0% + 5)
Input Impedance
: About 10 MΩ
Overload Protection
: 1000V DC / AC rms
Note:
The 1000V range is specified from 20% to 100% of range.
AC Voltage
Range
Resolution
Accuracy
6 V
0.001 V
± (1.0% + 5)
60 V
0.01 V
600 V
0.1 V
750 V
1 V
± (1.2% + 5)
Input Impedance
: About 10 MΩ
Frequency Range
: 40 Hz – 400 Hz
Response:
Average, calibrated in rms of sine wave
Overload Protection
: 1000V DC / AC rms
Note:
The 750V range is specified from 20% to 100% of range.
DC Current
Range
Resolution
Accuracy
60 mA
0.01 mA
± (1.0% + 7)
600 mA
0.1 mA
6 A
1 mA
± (1.5% + 7)
20 A
10 mA
Overload Protection:
Protection for
V
Ω
mAH
terminal inputs: Fuse, 630mA / 1000V,
Protection for
20A
terminal inputs: Fuse, 20A / 1000V, Fast
Max. Allowable Input Current
: 20A (For measurements >2A: measurement duration < 10 sec., and
interval> 15 minutes)
Note
: The 20A range is specified from 20% to 100% of range.
AC Current
Range
Resolution
Accuracy
60 mA
0.01 mA
± (1.8% + 10)
600 mA
0.1 mA
6 A
1 mA
± (2.5% + 10)
20 A
10 mA
Overload Protection:
Protection for
VΩmAHz
terminal inputs: Fuse, 630mAl1000V, Fast
Protection for
20A
terminal inputs: Fuse, 20Al1000V, Fast
Max. Allowable Input Current
: 20A (For measurements> 2A: measurement duration <10 sec, and
interval >15 minutes)
Frequency Range
: 40Hz - 400Hz
Response:
Average, calibrated in rms of sine wave
Note:
The 20A range is specified from 20% to 100% of range.