Diagnostics on the device
The unit does not power on
a) Have you pressed the On/Off switch for more than 2 seconds?
b) Check the batteries. Were they the right way used?
c) Replace the batteries with fresh batteries. Please try again.
The probe LED don not light after 60 seconds
a) Check whether the probe has a good contact. Should not change the condition of the unit, contact
your dealer.
Important notes
1. If the device gets contact with silicone vapors, the sensor will be dirty. Avoid the use of the device
where silicone vapors are present, where cleanser are used and where rubber is processed.
2. Avoid environments with a high concentration of corrosive gases such as H2S, SOx, CI2, HCI,
etc. The device could otherwise corrode and the main line and the heating unit of the device could
be damaged.
3. Measurement errors can occur when the probe is contaminated with alkaline metals or salt water
vapor comes into contact.
4. Measurement errors can be caused by contact with water (spray).
5. If probe comes in contact with water, the measurement results be distorted.
6. The device works best at an ambient air with an oxygen content of 21%. The device does not
work in an environment without oxygen or air with less than a 21%.
7. Slight condensation of the device if used in enclosed spaces should not be a problem. However, if
condensation liquid collecting on the probe, the measurement results would be distorted.
8. With a longer exposure of the device with high-concentration gases - even if the device is
switched off - in future the measurement can be distorted.
9. If the device is switched off for a long time, the probe could produce slight measurement
inaccuracies (whichever is how the device has been stored). So the device should be stored in a
sealed bag / bag of clean air.
10. Regardless of the power supply, the device will severely suffer if the probe extreme conditions
such as humidity, high temperature, or prolonged contamination with gases for a long period will
be suspended.
NOTE:
If not using the meter for long time, the device needs a longer warm-up phase. You can switch
it on from time to time, to avoid a long warm-up phase.